ATTACHING THE LINE CORD

 

 

BEFORE CONNECTING CONTROLLER TO AC POWER SUPPLY CHECK THE 
SETTING OF THE VOLTAGE SELECTOR MODULE. 
 
This module is on the back panel, bottom left. There are four settings. Be sure selector is set 
for your power requirements. A red indicator should be visible beside the correct voltage in 
the list printed on the fuse cover. If the red indicator does not mark the correct voltage, the 
card inside the fuse case must be pulled out and reinserted with the proper orientation.  

 
Please note: The controller can be set to 120V or 240V.However the pump and tank unit 
does not have voltage setting adjustments. Use PTU-110A for 110/120V and PTU-220 for 
220/240V operation. 
 
2.4 ATTACHING 

WATER 

TUBES 

 
The Pump and Tank Unit is supplied with two 
water tubes. All connections are self-sealing 
when disconnected to prevent water spills. 
To connect, push in firmly. Squeeze to 
disconnect.                                                                                                          
 
Connect one of the tubes from the OUTLET of pump to water INLET on the back panel of 
the controller. 
 
Connect the second tube from RETURN of pump to water OUTLET on the controller. 
 
 
  

 
 
 

 
 
 
 
 
 
 
 
 
 

 
 
2.5 

CONNECTING THE THERMAL PROBE 

 

 

Summary of Contents for NTE-2A

Page 1: ...t Supplied 2 2 Water Pump and Tank Unit 2 3 Attaching AC Line Cords 2 4 Attaching Water Tubes 2 5 Connecting the Thermal Probe 2 6 Connection to AC and Switch On 3 0 OPERATING INSTRUCTIONS 3 1 Adjusting Base Temperature of Controller 3 2 Measuring Absolute Temperature 3 3 Warning Light 4 0 SPECIFICATIONS 5 0 WARRANTY AND SERVICE W Shared docs Operating manuals EB NTE 2A Rev 1 01 25 07 ECO 10292 ...

Page 2: ...nd below room temperature The temperature of the tip of probe can be adjusted using the SET DIFFERENTIAL TEMP switch of the controller which maintains the temperature at the set point The digital readout on the controller continuously indicates the actual tip temperature of the probe For stable operation the thermoelectric module requires a trickle of cooling water This is supplied from the Pump a...

Page 3: ...lowing when the controller is turned on Check water level periodically during operation to ensure that water level is within three inches of the lip of the container To prevent spills all water connections are automatically self sealing when disconnected The pump in the PTU 110A Pump and Tank Unit is self priming When first started it will be noisy for five or ten seconds This is normal After the ...

Page 4: ...ut and reinserted with the proper orientation Please note The controller can be set to 120V or 240V However the pump and tank unit does not have voltage setting adjustments Use PTU 110A for 110 120V and PTU 220 for 220 240V operation 2 4 ATTACHING WATER TUBES The Pump and Tank Unit is supplied with two water tubes All connections are self sealing when disconnected to prevent water spills To connec...

Page 5: ...ion To protect the tip during shipment and when not in use a clear protective sheath is provided with the Thermal Probe The protective sleeve has a bayonet type locking feature and can be left in place on the probe shaft To expose the tip simply twist the sheath until the retaining screw is in the radial slot and slide the sheath back or forward along the probe shaft Twist it again to lock it in p...

Page 6: ...de exact setting All further temperature settings can be made using the digital control settings switches on the front panel These settings allow the temperature of the stage to be incremented or decremented in 5 1 degree or 1 10 steps from the base temperature 3 2 MEASURING ABSOLUTE TEMPERATURE The digital display of the controller may be used as a readout for an external sensor Type T copper con...

Page 7: ... of set point Safety shutdown with warning lamp in case of fault condition such as thermocouple sensor breakage lack of cooling water or electronics failure Self sealing water connectors Auxiliary AC switched output THERMAL PROBE SPECIFICATIONS SIZE 6 5 long x 5 8 dia Total length with cable 6 feet approx WEIGHT 0 5 lb TEMPERATURE CONTROLLED 0 5 diameter SURFACE AREA RANGE 0 45 C Note Sustained ex...

Page 8: ... this instrument call 973 779 5577 or e mail physitemp aol com In the event that any part of this system is to be returned for repair or recalibration please pack it with care and send it prepaid via Federal Express to Service Department PHYSITEMP INSTRUMENTS INC 154 Huron Avenue Clifton NJ 07013 U S A Please include with the instrument 1 A note describing any problems encountered 2 The name and t...

Reviews: