LIFEPAK 20/20e Defibrillator/Monitor
Replacement Procedures
Boardstack
(continued)
Boardstack Removal
(continued)
5. Lift the boardstack assembly out of its track and tilt it forward to make the
lower connections accessible.
6. Disconnect the W09 26-pin cable from the A04 Therapy PCB at J16 by
releasing the outer tabs.
7. Disconnect the W01 Therapy Connector Assembly from the A04 Therapy
PCB at J14.
8. Disconnect the W07 Capacitor Discharge Cable from the A04 Therapy PCB
at J2.
9. Disconnect the therapy connector cable at J13.
10. Turn the device so that the ECG and therapy connectors are in view.
11. Disconnect the W03 IrDA flex cable from the A01 System PCB at J8 by
removing the screw (255) using a 3/32 allen driver.
(Continued on next page)
9-73