LIFEPAK 15 Monitor/Defibrillator
Service Manual
Troubleshooting
Device Log
109
5
Device Log Entries
The Device Log includes the data listed in
.
Table 5.2—Device Log Entries
Data
Description
Fault Messages
Records YES or NO to indicate whether there are any error codes stored in
the Service Log (see
Processing Service Log Codes (p. 115)
Power Cycle Count
Number of times the device has been turned on
Pacing Count
Total pacing pulses delivered by the device
Shock Count
Total times the device defibrillation capacitor has been charged
Power On Time
Total device power-on time
Printer On Time
Total printer running time
SpO2 Operating Time
Total SpO2 running time
CO2 Operating Time
Total CO2 running time
NIBP Inflation Cycles
Total number of inflation cycles