LIFEPAK 15 Monitor/Defibrillator
Service Manual
Troubleshooting
Processing Service Log Codes
115
5
Processing Service Log Codes
When an internal program or process fails to execute properly, an error code is written into the device
(for
example, 500e), and the front panel Service LED (see
illuminates.
To process error codes:
1.
Review the error code(s) by displaying the
.
2.
Clear the Service Log (
Clearing the Service Log (p. 114)
), and then turn the device OFF. The CLEAR LOG action will record the
error(s) on the 100-mm printer.
3.
Complete the Performance Inspection Procedures (PIP).
~ If the PIP completes successfully, the device may be returned to regular use. The error code(s) may have been related
to EMI or ESD.
~ If the Service LED illuminates at any time during the PIP, stop the PIP and compare your PIP failure with the
. Continue with step 4.
4.
Locate the specific corrective action for a Service Log error code as follows:
a.
Display the Service Log to view the error code(s).
b.
Review the
Service Log Code Categories (p. 119)
for general information.
c.
Click the appropriate link in the Initial Digit column and locate your specific error code in the table.
d.
Click the link(s) in the Corrective Action column to view the corresponding corrective action.
e.
Service the device based on these inputs, and then repeat the Performance Inspection Procedures (PIP).
5.
For persistent Service Log codes, contact your local Physio-Control service or sales representative.