LIFEPAK 15 Monitor/Defibrillator
Service Manual
Troubleshooting
Troubleshooting Chart
99
5
Troubleshooting Chart
NOTE:
Corrective actions are listed in sequential order according to what is most likely to correct the observed symptom.
ge
Table 5.1— Corrective Actions
Area
Observed Symptom
Suggested Corrective Action
Physical Inspection
Loose or broken hardware
Locate and tighten or replace loose items.
Locate and replace broken components.
Evidence of dirt, fluids, or foreign
objects
Perform
External Cleaning Procedure (p. 154)
Damaged keypad or label
Possible A09 Printer Control Keypad failure.
Possible A10 Main Keypad failure.
Replace Bezel Label (158).
Replace Product Identification Label (162).
Replace Explosion/Hazard Label (164).
Damaged battery pin(s)
Possible battery pin(s) failure.
Power On/Self Test
No power ON
Install fully charged, properly maintained batteries.
Possible loose or broken battery pin(s).
Possible W04 System/Interface PCB cable connection issue.
Possible A03 Power PCB failure.
Service LED remains ON
See
Processing Service Log Codes (p. 115)
MAINTENANCE DUE
message
appears
See
Setting/Resetting the Maintenance Prompt Interval (p. 150)
.