![Photon Systems Instruments AquaPen-C AP 110-C Instruction Manual Download Page 14](http://html1.mh-extra.com/html/photon-systems-instruments/aquapen-c-ap-110-c/aquapen-c-ap-110-c_instruction-manual_1553565014.webp)
Page | 14
5
G
ETTING STARTED
For more detailed information on steps of AquaPen operation please refer to chapter
Error! Reference source not found.
.
The device is powered by built in Li-Ion battery. Ensure that the battery if fully charged by plugging it into a PC via USB cable or the AC
outlet via USB adapter (not included) and the cable.
The AquaPen is controlled using two buttons:
•
Use the
MENU
key to scroll through sequential menu options shown on the digital display. And to turn the device off (hold for 3
sec).
•
Use the
SET
key to turn the device on (hold for 3 sec) and select a menu option based on cursor (>) position.
5.1
M
EASUREMENTS BASED ON FLUORESCENCE
5.1.1
P
ULSES DESCRIPTION AND SETTING
Flash pulse
This function serves for setting of measuring pulses intensity. The measuring pulses are weak light pulses, which are able to induce the
minimal
chlorophyll
fluorescence
(F
0
or
F
t
).
It
takes
only
30
µs
and
the
maximum
intensity
is
3,000 µmol.m
-2
.s
-1
. It means 30 µs * 3,000 µmol.m
-2
.s
-1
= 0.09 µmol.m
-2
per pulse is the maximal intensity of the flash pulse.
Super pulse
This function serves for setting intensity of the saturating pulse. Saturating light pulse is able to induce maximum chlorophyll fluorescence
(F
m
). 100% of intensity represents approximately 3,000 µmol.m
-2
.s
-1
.
Actinic pulse
This function serves for setting intensity of actinic light. It is the ambient light in which the algae are growing. 100% of intensity equals
approximately 1,000 µmol.m
-2
.s
-1
.
Pulses used in predefined protocols:
Measurements based on fluorescence
Used pulses
Ft
Flash pulse
QY
Flash pulse, Super pulse
OJIP
Super pulse
NPQ protocols
Flash pulse, Super pulse, Actinic pulse
Light Curves
Flash pulse, Super pulse (Actinic pulse is preset)
Summary of Contents for AquaPen-C AP 110-C
Page 1: ......
Page 9: ...Page 9 3 DEVICE DESCRIPTION Fig 1 Device description...
Page 13: ...Page 13 Fig 3 AquaPen C AP110 C Fig 4 AquaPen P AP 110 P...
Page 22: ...Page 22 Fig 8 NPQ Protocol...
Page 24: ...Page 24 Fig 9 LC1 Protocol...
Page 25: ...Page 25 Fig 10 LC2 Protocol...
Page 29: ...Page 29...
Page 30: ...Page 30...
Page 31: ...Page 31...
Page 32: ...Page 32...
Page 33: ...Page 33...
Page 34: ...Page 34...
Page 35: ...Page 35...
Page 47: ...Page 47 Fig 26 Data export...
Page 51: ...Page 51 Fig 34 GPS coordinates...