Additional information
MC10 Control Unit
The following standards and norms were used in the versions with at least IPX4 and higher - in
according to:
-
EN 60601-2-52, IEC 60601-2-52, (Particular requirements for the safety and essential
performance of medical beds)
IEC/EN 60601-2-52, Section 201.6.2
Protection against electrical shock: Protection class II
IEC/EN 60601-2-52, Section
201.7.6.3
Control panel symbols (depending on model, customer
requirements)
IEC/EN 60601-2-52, Section
201.8.11.3.2
EPR-Power supply lead: only
≥ 2.5 m length
IEC/EN 60601-2-52, Section
201.9.2.2.5
Continuous operations: Control unit only with button
IEC/EN 60601-2-52, Section
201.9.2.3.1
Unintentional movement:
Prevented by means of a locking mechanism (such as
Control box, Supervisor, IPROXX
®
SE, IPROXX
®
, or
Meditouch)
IEC/EN 60601-2-52, Section
201.9.6.2.1
Noise level:
≤65dB(A)
IEC/EN 60601-2-52, Section
201.11.1.1
Temperature
IEC/EN 60601-2-52, Section
201.11.6.5.101
Protection against water ingress: only for
≥ IPX4
IEC/EN 60601-2-52, Section
201.11.8
Loss of power:
e.g. use of a battery, depending on customer
requirements
IEC/EN 60601-2-52, Section
201.13.1.4
Special mechanical risks, depending on customer
requirements:
Prevented by means of a locking mechanism (such as
Control box, Supervisor, IPROXX
®
SE, IPROXX
®
, or
Meditouch)
IEC/EN 60601-2-52, Section
201.15.3.4.1
Mechanical strength
– handset (e.g. IPROXX
®
)
IEC/EN 60601-2-52, Section
201.15.4.4
Displays: Ready indicator is not required
IEC/EN 60601-2-52, Section
BB.3.3.3
Dimensions (depending on version and customer
requirements)
– handset (e.g. IPROXX
®
)
IEC/EN 60601-2-52, Section
BB.3.4.1
Operational forces
– handset (e.g. IPROXX
®
)