16
20 kHz
10 Hz
± (0.01 % of the measured value + 1 digits)
> 1.5 < 5 V
eff
200 kHz
100 Hz
± (0.01 % of the measured value + 1 digits)
> 1.5 < 5 V
eff
2 MHz
1 kHz
± (0.01 % of the measured value + 1 digits)
> 2 < 5 V
eff
20 MHz
10 kHz
± (0.01 % of the measured value + 1 digits)
> 2 < 5 V
eff
7.9 Temperature ranges °C
( teStFOX M-1)
temperature measurements ( teStFOX M-1) are possible by means of the
enclosed temperature measuring adapter only.
Overload protection for temperature measurements: 600 V
eff
Measuring range
Resolution
Measuring accuracy
-20 °C ~ 0 °C
1 °C
± (2 % + 4 °C)
1 °C ~ 100 °C
1 °C
± (1 % + 3 °C)
101 °C ~ 500 °C
1 °C
± (2 % + 3 °C)
501 °C ~ 800 °C
1 °C
± (3 % + 2 °C)
7.10 Temperature ranges °F
( teStFOX M-1)
temperature measurements ( teStFOX M-1) are possible by means of the
enclosed temperature measuring adapter only.
Overload protection for temperature measurements: 600 V
eff
Measuring range
Resolution
Measuring accuracy
-4 °F ~ 32 °F
1 °F
± (2 % + 8 °F)
33 °F ~ 212 °F
1 °F
± (1 % + 6 °F)
213 °F ~ 932 °F
1 °F
± (2 % + 6 °F)
933 °F ~ 1472 °F
1 °F
± (3 % + 4 °F)
8. Measuring by means of the TESTFOX M/ M-1
8.1 Preparing the measurement
Operate and store the TESTFOX M/ M-1 at the specified storage and operating
temperatures only!
Do not permanently expose the device to sunlight.
- Check stated nominal voltage and nominal current on the safety measuring
lines. Nominal voltage and current of the enclosed safety measuring lines
comply with the respective values of the teStFOX M/ M-1.
- Check insulation of the safety measuring lines. if the insulation is damaged,
the safety measuring lines must be replaced immediately.
- Check the safety measuring lines for continuity. if the conductor in the
safety measuring line is interrupted, replace the safety measuring lines
immediately.
- Before selecting another function by means of the rotary switch
8
,
disconnect the safety measuring lines from the measuring point.
- Strong sources of interference in the vicinity of the teStFOX M/ M-1 might
involve unstable readings and measuring errors.
8.2 Voltage and current measurement
Do not exceed the maximum permitted voltage with respect to
earth potential! Electrical danger!
the highest voltage which may be applied to the
- COM jack
J
- jack for
V, Ω,
, Hz
9
- jack for the 10 A range
K
( teStFOX M-1)
of the teStFOX M/ M-1 against ground is 1000 V.
Electrical danger!
Maximum circuit voltage for a measured current of 500 V! In
case of fuse triggering at values above 500 V, damages of the
device might be involved. A damaged device might represent
an electrical hazard!
8.2.1 Voltage measurement
- Select the desired function (V AC) or (V DC) by means of the rotary switch
8
of the teStFOX M/ M-1.
- Connect the black safety measuring line to the COM jack
J
of the
teStFOX M/ M-1.
10/ 2008
TESTFOX M/ M–1