3.3.1.

With timer control

Operation, using a timer, is as follows:

1 . Check that generator  kV and  mA settings are correct.

2.

Determine the exposure time, and set the timer for the required window shutter. The clear plastic cover is
rotated until the black pointer is 

 

set to the required time  (0 to  6),  while the centre knob is rotated in a

clockwise direction until the required multiplication factor (x1, x10 : seconds, minutes or hours) is against

 

the arrow.

When timing commences the red pointer will gradually return to zero, then the timer microswitches return

to the normal position (off), the timer and the red pointer returns to the position set by the black

pointer.

3. Set the three   -  position SHUTTERS switch to select the required window shutter, e.g. when window 3 is to

be used the right-hand switch is set to 3; and the time must be set by the right-hand timer.

4.

Open the shutter by operating the lowest SHUTTERS push-button simultaneously with the required
SHUTTERS I push-button. Whent he shutter is open the X-ray beam passes to the sample in the camera,

the red lamp on the tube shield (below the filter/shutter disc) is lit, the red flags will appear in the holes

on either side of the tube shield, and the timer operation will commence.

5. When the preset time has elapsed the shutter is closed, and the whole instrument is switched off unless a

shutter is open at another tube shield position. X-rays will be generated as long as one or more shutters
are open.

NOTE:

The red warning lamps are fail safe, and the shutters will close if a bulb filament is broken when a

shutter is open.

3.3.2.

Without timers

When a timer is not used, or is not fitted, proceed as follows:

1 .

Check that the generator  kV and  mA settings are correct.

2. Ensure that the 3   - position switches are in the mid-position.

3.

Open the shutter by operating the lowest SHUTTERS push-button simultaneously with the required
SHUTTERS I push-button. The shutter will open and the X-ray beam passes to the sample in the camera,

the  red  lamp on the tube shield is lit, and the red flags will appear in the holes on either side of the tube shield.

4.

When the exposure time has elapsed, operate the SHUTTER 0 push-button. The shutter will  close,

and the generator will switch off unless a shutter is open at another tube shield position. X-rays will be

generated as long as one or more shutters are open.

3.4.

SWITCHING OFF A MOTOR WHEN THE SHUTTER CLOSES

Cameras can be fitted with motors, either for moving film, or a crystal, or for spinning the sample.  If required.

the motor operation can be switched off when the timer reaches its zero position, and it will remain off
until the timer is reset (if one or more shutters are still open). The two outside a.c. mains sockets on the rear

panel are used for this purpose, and it is necessary for a jumper wire to be removed between terminals 25 and

26 on the rear of the timer, refer to the Service Part.

The mains connector for the motor used in the tube shield position 1 or 2, must be plugged into the mains
socket immediately behind, i.e. adjacent to the left-hand side panel. A motor used for positions 3 or 4 will be

connected to the left-hand socket on the rear panel (viewed from the rear).

The procedure for analysis is as described in paragraph  3.3.1.,  but now the camera motor will start at the moment

that the Starting relay operates, when GENERATOR I is depressed.
When the shutter is opened the timer will start, and at the end of the preset time when the shutter closes,

the motor will also stop. If another shutter is still open the equipment will not switch off, the timer is not reset
and there is no power to the motor.

To reset the timer and start the camera motor, place the three-position switch in its central position. If all
shutters are closed after the preset time, the equipment switches off and the timer resets.

Summary of Contents for PW 1720

Page 1: ...PHILIPS Table top X ray Generator PW 1720 Instruction manual ...

Page 2: ...sis by loading the sample film etc or positioning the goniometer or other device see also paragraph 3 4 l Momentarily operate the GENERATOR I push button and when the safety circuit is complete the X RAYS ON lamp s will light to show that X rays are being generated The lamp in the push button is also lit l Increase the kV control to the required value then adjust the mA control until the desired X...

Page 3: ...osure time of the film to X rays by using a PW 1722 timer Two timers can be fitted the left hand timer will control the shutter of either window 1 or window 2 and the other timer the shutter of window 3 or 4 This paragraph describes the automatic and manual control of the shutter operation When the camera is installed on its bracket in front of the required exit window of the tube shield open the ...

Page 4: ...operating the lowest SHUTTERS push button simultaneously with the required SHUTTERS I push button The shutter will open and the X ray beam passes to the sample in the camera the red lamp on the tube shield is lit and the red flags will appear in the holes on either side of the tube shield 4 When the exposure time has elapsed operate the SHUTTER 0 push button The shutter will close and the generato...

Page 5: ... The filament has a rectangular shape but the size depends on the type of X ray tube Focal sizes are Fine focus 0 4 x 8 mm Normal focus 1 x 10 mm Broad focus tube 2x 12 mm Two windows in the metal body are parallel with the long side of the focus the two other windows are at right angler to it The windows are at the same height with respect to the anode plane and they allow beam take off angler be...

Page 6: ... handling the X ray tube take care to hold by the metal base and DO NOT TOUCH THE BERYLLIUM WINDOWS These windows are thin and can be poisonous to the skin Do not remove the X ray tube if generator operation has been switched off because of a failure safety circuit actuated see Service section 5 Select the new tube and insert into the tube shield after first checking the condition of the 0 rings f...

Page 7: ... generator i e so that the X ray beams make angles of either 90 or 45 or 0 to the front of the generator However the rotation of the tube shield must not be more than 90 in either direction from its original position Since the line and point focus beams are 90 apart it is possible to place the tube shield with either beam in any position depending on the use of the generator or personal preference...