When the shutter is closed the generator will switch off, unless another shutter is still  open  in one of the

other tube shield positions. Alternatively. the generator can be  switched off manually by operating the

GENERATOR 0 push-button. All lamps will extinguish. Before the generator is switched off the power

to the X-ray tube should be reduced, by first turning the  mA control to minimum, and then the  kV  control.
Any generation of HT will be inhibited if the shutter does not fully close after a close shutter command.

Fig. 3.1.

Opening  the  window  shutter.

3.3.

SHUTTER CONTROL

For camera work it is usual  to control the exposure time of the film to X-rays by using a PW 1722 timer. Two

timers can be fitted: the  left-hand  timer will control the shutter of either window 1 or window 2, and the
other timer the shutter of window 3 or 4. This paragraph describes the automatic and manual control of the

shutter operation.
When the camera is installed on  its  bracket in front of the required exit window of the tube shield, open the
shutter disc and push the camera forward until the collimator enters the hole in the disc. Ensure that the

microswitch button on the tube shield is actuated by the pin or arm of the  camera.

NOTE:

The space between the front of the collimator and the exit opening of the tube shield should be kept

as  small  as possible to ensure  that radiation does  not escape. Take care  that the filter disc is free

to   move,  and is not damaged.

Summary of Contents for PW 1720

Page 1: ...PHILIPS Table top X ray Generator PW 1720 Instruction manual ...

Page 2: ...sis by loading the sample film etc or positioning the goniometer or other device see also paragraph 3 4 l Momentarily operate the GENERATOR I push button and when the safety circuit is complete the X RAYS ON lamp s will light to show that X rays are being generated The lamp in the push button is also lit l Increase the kV control to the required value then adjust the mA control until the desired X...

Page 3: ...osure time of the film to X rays by using a PW 1722 timer Two timers can be fitted the left hand timer will control the shutter of either window 1 or window 2 and the other timer the shutter of window 3 or 4 This paragraph describes the automatic and manual control of the shutter operation When the camera is installed on its bracket in front of the required exit window of the tube shield open the ...

Page 4: ...operating the lowest SHUTTERS push button simultaneously with the required SHUTTERS I push button The shutter will open and the X ray beam passes to the sample in the camera the red lamp on the tube shield is lit and the red flags will appear in the holes on either side of the tube shield 4 When the exposure time has elapsed operate the SHUTTER 0 push button The shutter will close and the generato...

Page 5: ... The filament has a rectangular shape but the size depends on the type of X ray tube Focal sizes are Fine focus 0 4 x 8 mm Normal focus 1 x 10 mm Broad focus tube 2x 12 mm Two windows in the metal body are parallel with the long side of the focus the two other windows are at right angler to it The windows are at the same height with respect to the anode plane and they allow beam take off angler be...

Page 6: ... handling the X ray tube take care to hold by the metal base and DO NOT TOUCH THE BERYLLIUM WINDOWS These windows are thin and can be poisonous to the skin Do not remove the X ray tube if generator operation has been switched off because of a failure safety circuit actuated see Service section 5 Select the new tube and insert into the tube shield after first checking the condition of the 0 rings f...

Page 7: ... generator i e so that the X ray beams make angles of either 90 or 45 or 0 to the front of the generator However the rotation of the tube shield must not be more than 90 in either direction from its original position Since the line and point focus beams are 90 apart it is possible to place the tube shield with either beam in any position depending on the use of the generator or personal preference...

Page 8: ...3 7 ...

Reviews: