3. OPERATION

This paragraph describes the method of operating the PW 1720 and the options, such as timers etc., but does

not include operating instructions for accessory devices such as cameras and vertical goniometers.

The initial installation of the PW 1720 is described in the Service part of the manual.

3.1.

INITIAL CHECKS

Before switching on the generator, check the following:

- Check that HT tank transport fixing screw is removed, see Service installation section.

- Cooling water for the X-ray tube is turned on.

- The tube shield is fixed in the required position ( 0 ° ,  4 5 ° , or 90°).

- The X-ray tube installed is suitable for the analysis program.

- The kV  and mA  controls are in minimum position (10 kV and 5 mA),  also refer to paragraph 3.6.

- The PW 1161 X-RAYS ON lamp (if fitted) is correctly wired, and is placed in a conspicuous position.

- The timers (if required) are installed.

- The accessory units, required for the analysis program, are correctly fitted to the tube shield.

The white push-button must be depressed by the accessory, and the shutter disc is rotated to allow the

X-ray beam to pass.

- Set the filter disc to the desired position.

- Check that the a.c. mains stabiliser is correctly installed, when required.

3.2. 

SAMPLE ANALYSIS

The procedure for analysis will depend on the accessories being used, and the operation of these devices is

described separately in the respective manuals. They should be made ready for analysis by loading the sample,

film (etc.), or positioning the goniometer or other device -see also paragraph 3.4.

l

 Momentarily operate the GENERATOR I push-button, and when the safety circuit is complete the X-RAYS

ON lamp(s) will light to show that X-rays are being generated. The lamp in the push-button is also lit.

l

 Increase the kV  control to the required value, then adjust the mA control until the desired X-ray tube

current is indicated by the meter. Do not exceed the rated maximum power to the X-ray tube, i.e.

the setting of  kV and  mA must not exceed the values shown for X-ray tubes in Appendix 1. The ratings

shown in the “X-ray diffraction tubes” leaflet (9499 300 10311) are not applicable.

l

  Determine the exposure time, and set the timer if fitted - also refer to paragraph 3.3.

l

 Place the SHUTTERS three -  position switch to select the desired window shutter (with timer), or

infinite time (middle position).

l

 Open the shutter by simultaneously pressing the lowest SHUTTERS push-button, and the SHUTTERS I

push-button corresponding to the window shutter which is to be opened, refer to figure 3.1. The shutter
will remain open until the timer exposure time has elapsed, or until the SHUTTERS 0 push-button is

operated (when a timer is not fitted or used).

During the time that the shutter is opened, the X-ray beam is allowed into the camera or goniometer, the red

lamps on the tube shield below the shutter/filter disc are lit, and the red flags will appear in the side holes

of the cover over the tube shield mechanism

In the event of a failure of one of the lamp bulbs in the

tube shield, the shutter is closed immediately.

Summary of Contents for PW 1720

Page 1: ...PHILIPS Table top X ray Generator PW 1720 Instruction manual ...

Page 2: ...sis by loading the sample film etc or positioning the goniometer or other device see also paragraph 3 4 l Momentarily operate the GENERATOR I push button and when the safety circuit is complete the X RAYS ON lamp s will light to show that X rays are being generated The lamp in the push button is also lit l Increase the kV control to the required value then adjust the mA control until the desired X...

Page 3: ...osure time of the film to X rays by using a PW 1722 timer Two timers can be fitted the left hand timer will control the shutter of either window 1 or window 2 and the other timer the shutter of window 3 or 4 This paragraph describes the automatic and manual control of the shutter operation When the camera is installed on its bracket in front of the required exit window of the tube shield open the ...

Page 4: ...operating the lowest SHUTTERS push button simultaneously with the required SHUTTERS I push button The shutter will open and the X ray beam passes to the sample in the camera the red lamp on the tube shield is lit and the red flags will appear in the holes on either side of the tube shield 4 When the exposure time has elapsed operate the SHUTTER 0 push button The shutter will close and the generato...

Page 5: ... The filament has a rectangular shape but the size depends on the type of X ray tube Focal sizes are Fine focus 0 4 x 8 mm Normal focus 1 x 10 mm Broad focus tube 2x 12 mm Two windows in the metal body are parallel with the long side of the focus the two other windows are at right angler to it The windows are at the same height with respect to the anode plane and they allow beam take off angler be...

Page 6: ... handling the X ray tube take care to hold by the metal base and DO NOT TOUCH THE BERYLLIUM WINDOWS These windows are thin and can be poisonous to the skin Do not remove the X ray tube if generator operation has been switched off because of a failure safety circuit actuated see Service section 5 Select the new tube and insert into the tube shield after first checking the condition of the 0 rings f...

Page 7: ... generator i e so that the X ray beams make angles of either 90 or 45 or 0 to the front of the generator However the rotation of the tube shield must not be more than 90 in either direction from its original position Since the line and point focus beams are 90 apart it is possible to place the tube shield with either beam in any position depending on the use of the generator or personal preference...

Page 8: ...3 7 ...

Reviews: