Approved Transducers for Electromagnetic Compliance
The imaging transducers used with the system may affect its emissions. The transducers listed
in
“Clinical Applications and Transducers” on page 174
, when used with the system, have been
tested to comply with the Group 1, Class B emissions, as required by International Standard
CISPR 11. Use only those transducers.
WARNING
Using cables, transducers, and accessories other than those specified for use with the system
may result in increased emissions from, or decreased immunity of, the system.
NOTE
The use of an Intracardiac Echocardiography (ICE) Probe or an Transesophageal
Echocardiogram (TEE) probe results in Class A Radiated Emissions and should be used in a
hospitable environment only.
Approved Accessories for Electromagnetic Compliance
Accessories used with the system may affect its emissions. The accessories listed here, when
used with the system, have been tested to comply with the Group 1, Class B emissions as
required by International Standard CISPR 11. Use only the accessories listed here.
Electromagnetic Compatibility
Safety
EPIQ 7 User Manual 4535 617 25341
73
Summary of Contents for epiq 7
Page 4: ...4 EPIQ 7 User Manual 4535 617 25341 ...
Page 26: ...Read This First Recycling Reuse and Disposal 26 EPIQ 7 User Manual 4535 617 25341 ...
Page 94: ...DVD RW Drive System Overview System Components 94 EPIQ 7 User Manual 4535 617 25341 ...
Page 154: ...Customizing the System Custom Procedures 154 EPIQ 7 User Manual 4535 617 25341 ...
Page 172: ...Performing an Exam Ending an Exam 172 EPIQ 7 User Manual 4535 617 25341 ...
Page 298: ...System Maintenance For Assistance 298 EPIQ 7 User Manual 4535 617 25341 ...