10. Additional Notes
Maximum dynamic pressure variation, according to ISO 80601-2-70:2015:
Tube Type
10 BPM
15 BPM
20 BPM
12 (mm) tubing type
± 0.6 cmH
2
O ± 0.8 cmH
2
O ± 1.6 cmH
2
O
15 (mm) tubing type and 22 (mm) tubing type ± 0.7 cmH
2
O ± 0.7 cmH
2
O ± 1.0 cmH
2
O
Dynamic pressure accuracy has a measurement uncertainty of 3.6%.
Maximum Flow Rate (typical)
Tube Type
Flow
Test pressures (cmH
2
O)
4.0
8.0
12.0
16.0
20.0
12 (mm) tubing type Average flow at the patient
connection port (l/min)
90
119
112
106
99
15 (mm) tubing type Average flow at the patient
connection port (l/min)
77
115
112
105
106
22 (mm) tubing type Average flow at the patient
connection port (l/min)
80
121
127
121
109
Disposal
Separate collection for electrical and electronic equipment per EC Directive 2012/19/EU.
Dispose of this device in accordance with local regulations.
EMC information
Your unit has been designed to meet EMC standards throughout its service life without
additional maintenance. There is always an opportunity to relocate your DreamStation
Go Therapy Device within an environment that contains other devices with their own
unknown EMC behavior. If you believe your unit is affected by locating it closer to another
device, simply separate the devices to remove the condition.
Pressure and Flow Accuracy
The DreamStation Go Therapy Device is designed to perform within the pressure and
flowrate accuracies specified in the user manual. If you suspect that the pressure and/or
flow rate accuracy is affected by EMC interference remove power and relocate the device
to another area. If performance continues to be affected discontinue use and contact
your supplier.
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions – This device is
intended for use in the electromagnetic environment specified below. The user of this
device should make sure it is used in such an environment.