Problem

Possible

cause

Solution

 

You are using

medication. 

The variations in

blood pressure can

be greater if you

are using medica­

tion.

 

You per­

formed multi­

ple measure­

ments direct­

ly after each

other. 

Wait at least 3

minutes between

measurements.

This allows your

blood circulation to

recover.

My blood

pressure

measure­

ment from

the hospital

is different

from the

measure­

ment at

home. 

Multiple vari­

ables may

affect your

blood pres­

sure such as

the weath­

er, emotions

and exercise.

Pay attention when

you measure your

blood pressure at

home. Check for

instance:

If the cuff is not too

tight or too loose.

If the cuff is prop­

erly attached on

the wrist.

If you feel anxious

or stressed, try to

relax. Take a deep

breath 2-3 times

before you start a

measurement.

Advice: Rest for 5

minutes before you

measure your

blood pressure.

The result is

different

when I per­

form mea­

surements

on my right

wrist.

The blood

pressure

monitor is

suitable to be

used on both

wrists, but the

measurement

results on the

right wrist

and left wrist

will differ.

For a meaningful

comparison, try to

measure under

similar conditions

and measure on

the same wrist

every time. 

The blood

pressure

monitor

does not

work when I

press the

'on' button

The

rechargeable

battery is

empty.

Recharge the bat­

tery (see 'Charg­

ing').

The lock

switch is set

to the

'locked' pos­

tion.

Set the lock switch

to the 'unlocked'

position (Fig. 7).

The light of

the display

dims and a

battery

Lo

is showing

The battery is

low.

Charge the battery

(see 'Charging').

The display

shows Err

Communica­

tion error.

Check if the app is

on and try data

transmission again.

The display

shows E3

The cuff is

not properly

secured.

Refasten the cuff,

wait for 3 minutes

and then measure

again.

The display

shows E10

or E11

The device

detected

motion, talk­

ing or the

heart rate is

too weak

during the

measure­

ment.

Wait for 3

minutes and then

measure again. Do

not move during

measurement. 

The display

shows E20

The device

does not

detect the

heart rate.

Make sure the

device is in contact

with the skin.

Loosen the cloth­

ing on the arm and

measure again.

The display

show E21

The mea­

surement

failed.

Wait for 3 minutes

and then measure

again.

The display

shows EExx

A system

error

occurred.

Retake the mea­

surement. If the

problem per­

sists, contact the

Philips Consumer

Care Center in your

country.

Data trans­

mission or

pairing

failed.

Bluetooth is

off.

Turn on Bluetooth

on your mobile

device.

The Philips

HealthSuite

health app is

off.

Press the icon on

your mobile device

to activate the

health app.

The blood

pressure

monitor and

mobile

device are

more than 5

meters away

from each

other. 

Place your mobile

device closer to the

blood pressure

monitor. 

You selected

the wrong

profile on the

blood pres­

sure monitor.

Select the correct

user profile on the

blood pressure

monitor before

your measurement.

Otherwise the data

cannot be trans­

mitted to your app.

Repeat the mea­

surement with the

correct profile

selected

Summary of Contents for DL8765

Page 1: ...lijke Philips N V All rights reserved Philips Consumer Lifestyle BV Tussendiepen 4 9206AD Drachten Netherlands Fax 31 0 512594316 4222 100 4515 2 10 2015 1 2 2 3 4 DL8765 5 6 7 8 9 10 1 2 cm 0 4 0 8 11 12 13 14 15 16 17 18 19 20 75 recycled paper 75 papier recyclé ...

Page 2: ...ctly and to protect you and others from injury Below you find the meaning of the warning signs and symbols on the label and in the user manual Symbol for follow instructions for use This symbol means that the part of the device that comes into physical contact with the user also known as the applied part is of type BF Body Floating accord ing to IEC 60601 1 The applied part is the cuff Symbol for ...

Page 3: ... transfu sion or an arterio venous shunt A V shunt is present The temporary interference to blood flow by the blood pressure measurement could result in injury If you had a mastectomy breast amputation do not use this blood pressure monitor on the arm on the side of the mastectomy The inflat ing cuff can lead to pain trauma and further injury in the arm on the side of the mastectomy Consult your d...

Page 4: ...tery can no longer be recharged or when the device does not function properly anymore see Specifications please contact the Philips Consumer Care center in your coun try Keep the device away from fire and heat sources as the battery can overheat causing fire or bursting The battery could explode causing injury or death After charging remove the micro USB plug from the device and remove the USB plu...

Page 5: ...charging indications Battery symbol Battery charging indication Battery charging half full Battery charging almost full Battery fully charged Using the blood pressure monitor This tubeless device uses the oscillometric method to measure blood pressure and heart rate Before every measurement the unit establishes a zero point equivalent to the atmospheric pres sure Then it starts inflating the cuff ...

Page 6: ...re monitor is equipped with Blue tooth Smart You can receive your personal health data on a mobile device that is equipped with the Bluetooth Smart function Download the Philips HealthSuite health app from the App store or Google Play Use the search term Philips Health Suite health app The app is available for iOS 8 0 and Android 4 4 Note You can only use the Philips HealthSuite health app to comm...

Page 7: ... automatically selects the previous user To change the user profile press the user ID button and select a different user Fig 14 Make sure the correct user is selected so the measurement data is properly transmitted and stored Also a guest user can be selected The guest user is for performing a measurement on other persons without a user profile in the app Measurements performed when using the gues...

Page 8: ... for the separate collection of electrical and electronic products and rechargeable batteries Correct disposal helps prevent negative consequences for the environment and human health Removing the rechargeable battery Warning This procedure is irreversible You cannot use the device anymore after this pro cedure Note We strongly advise you to take your product to an official collection point or a P...

Page 9: ...ked position Fig 7 The light of the display dims and a battery symbol Lo is showing The battery is low Charge the battery see Charging The display shows Err Communica tion error Check if the app is on and try data transmission again The display shows E3 The cuff is not properly secured Refasten the cuff wait for 3 minutes and then measure again The display shows E10 or E11 The device detected moti...

Page 10: ...c environments EMC Guidance The Blood Pressure Monitor needs special pre cautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the accompanying documents Wireless communications equipment such as wireless home network devices mobile phones cordless telephones and their base stations walkie talkies can affect this equipment and should be ...

Page 11: ...e quency 50 60 Hz magnet ic field IEC 61000 4 8 3A m 3A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commer cial or hospital environment Note UT is the AC mains voltage prior to applica tion of the test level Guidance and manufacturer s declara tion electromagnetic immunity for ME equipment and ME systems that are not life supporting Guida...

Page 12: ...ation distances between portable and mobile RF communications equip ment and the device The device is intended for use in an electromag netic environment in which radiated RF distur bances are controlled The customer or the user of the device can help prevent electromagnetic inter ference by maintaining a minimum distance between portable and mobile RF communications equipment transmittters and th...

Reviews: