Emissions test Com
pliance
Electromagnetic en
vironment - gui
dance
RF emissions
CISPR 11
Group
2
The device must emit
electromagnetic ener
gy in order to perform
its intended function.
Nearby electronic
equipment may be af
fected.
RF emissions
CISPR 11
Class B
Harmonic emis
sions IEC
61000-3-2
Not ap
plicable
Voltage fluctuati
ons/flicker emis
sions IEC
61000-3-3
Not ap
plicable
Guidance and manufacturer's declara
tion – electromagnetic immunity – for
all ME equipment and ME systems
Guidance and manufacturer’s declaration – electro
magnetic immunity
The device is intended for use in the electromagne
tic environment specified below. The customer or
the user of the device should assure that it is used
in such an environment.
Immuni
ty test
IEC
60601
test le
vel
Com
pliance
level
Electromagnetic
environment -
guidance
Electro
static
dischar
ge (ESD)
IEC
61000-4
-2
±6 kV
contact
±8 kV air
±6 kV
contact
±8 kV
air
Floors should be
wood, concrete or
ceramic tile. If
floors are covered
with synthetic ma
terial, the relative
humidity should
be at least 30%.
Power
frequen
cy
(50/60H
z) ma
gnetic
field IEC
61000-4
-8
3A/m
3A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in
a typical commerci
al or hospital envi
ronment.
Guidance and manufacturer's declara
tion – electromagnetic immunity –for
ME equipment and ME systems that
are not life supporting
Guidance and manufacturer’s declaration – electro
magnetic immunity
The device is intended for use in the electromagne
tic environment specified below. The customer or
the user of the device should assure that it is used
in such an environment.
IMMUNI
TY test
IEC
606
01
TES
T LE
VEL
Com
plia
nce
le
vel
Electromagnetic envi
ronment - guidance
Conduc
ted RF
IEC
61000-4-
6
3
Vrms
150
kHz
to
80
MHz
3
Vrms
Radiated
RF
IEC
61000-4-
3
3
V/m
80
MHz
to
2.5
GHz
3
V/m
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the device, in
cluding cables, than
the recommended se
paration distance calcu
lated from the equati
on applicable to the
frequency of the trans
mitter.
Recommended separa
tion distance:
d = 1.167
ÖP
d = 1.167
ÖP
80 MHz to
2.5 GHz
d = 2.333
ÖP
800 MHz
to 2.5 GHz
where
P
is the maxi
mum output power ra
ting of the transmitter
in watts (W) according
to the transmitter ma
nufacturer and
d
is the
recommended separati
on distance in metres
(m).
Field strengths from fi
xed RF transmitters, as
determined by an elec
tromagnetic site survey
(a), should be less than
the compliance level in
each frequency range
(b).
Interference may occur
in the vicinity of equip
ment marked with the
following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher fre
quency range applies.
NOTE 2 These guidelines may not apply in all situa
tions. Electromagnetic propagation is affected by
absorption and reflection from structures, objects
and people.
(a) Field strengths from fixed transmitters, such as
base stations for radio (cellular/cordless) telepho
nes and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess
the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should
be considered. If the measured field strength in the
location in which the device is used exceeds the
applicable RF compliance level above, the device
should be observed to verify normal operation. If
abnormal performance is observed, additional mea