This symbol on the device means: protec
ted against access to hazardous parts with
a finger and against vertically falling water
drops when tilted up to 15 degrees.
–20ºC
60ºC
Indicates the storage and transportation
temperature limits to which the medical
device can be safely exposed: -20°C to
60°C.
Symbol for the 2 year Philips guarantee.
The Green Dot ('Der Grüne Punkt' in Ger
man) is the license symbol of a European
network of industry-funded systems for
recycling the packaging materials of consu
mer goods.
Electromagnetic emissions and
immunity
The device is approved according to EMC safety
standard EN 60601-1-2. It is designed to be used in
typical domestic environments.
EMC Guidance
- The ear thermometer needs special precautions
regarding EMC and needs to be installed and
put into service according to the EMC informati
on provided in this document.
- Wireless communications equipment such as
wireless home network devices, mobile phones,
cordless telephones and their base stations, wal
kie-talkies can affect this equipment and
should be kept at least a distance d = 3.3 m
away from the equipment.
Note: As indicated in IEC 60601-1-2:2007 for ME
equipment, a typical cell phone with a maximum
output power of 2 W yields d = 3.3 m at an immuni
ty level of 3V/m.
Guidance and manufacturer's declara
tion – electromagnetic emissions - for
all ME equipment and ME systems
Guidance and manufacturer’s declaration – electro
magnetic emissions
The device is intended for use in the electromagne
tic environment specified below. The customer or
the user of the device should assure that it is used
in such an environment.
Emissions test Compli
ance
Electromagnetic
environment - guid
ance
RF emissions
CISPR 11
Group
2
The device must emit
electromagnetic ener
gy in order to perform
its intended function.
Nearby electronic
equipment may be
affected.
RF emissions
CISPR 11
Class B
Harmonic emis
sions IEC
61000-3-2
Not
applica
ble
Voltage fluctuati
ons/flicker emis
sions IEC
61000-3-3
Not
applica
ble
Guidance and manufacturer's declara
tion – electromagnetic immunity – for
all ME equipment and ME systems
Guidance and manufacturer’s declaration – electro
magnetic immunity
The device is intended for use in the electromagne
tic environment specified below. The customer or
the user of the device should assure that it is used
in such an environment.
Immuni
ty test
IEC
60601
test
level
Compli
ance
level
Electromagnetic
environment -
guidance
Electro
static
dischar
ge (ESD)
IEC
61000-4
-2
±6 kV
contact
±8 kV air
±6 kV
contact
±8 kV
air
Floors should be
wood, concrete or
ceramic tile. If
floors are covered
with synthetic
material, the relati
ve humidity
should be at least
30%.
Power
frequen
cy
(50/60H
z) mag
netic
field IEC
61000-4
-8
3A/m
3A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in
a typical commerci
al or hospital envi
ronment.
Guidance and manufacturer's declara
tion – electromagnetic immunity –for
ME equipment and ME systems that
are not life supporting
Guidance and manufacturer’s declaration – electro
magnetic immunity
The device is intended for use in the electromagne
tic environment specified below. The customer or
the user of the device should assure that it is used
in such an environment.
IMMUNI
TY test
IEC
606
01
TES
T
LEV
EL
Com
pli
an
ce
leve
l
Electromagnetic envi
ronment - guidance