Meridian 2 97
Storage
- Temperature:
-20 to 60 degree C
-
Humidity:
95%
max
- Altitude:
0 -12192ft
- Air pressure:
300-1100 mBAR
Note
: recommend at 5 to 35
°
C, Humidity less than 60%.
6.2 Transportation tests
Refer to 5.15.2
6.3 Display disturbances from external environment
According to IEC 801-2 for ESD disturbances
6.4 Display disturbances to external environment
7. Reliability
7.1 Mean Time between Failures
System MTBF (Including the LCD panel and CCFL): 50,000 hrs
8. Quality Assurance Requirements
8.1 Acceptance test
According to MIL-STD-1916D Control II level
AQL: NA
(Please also refer to annual quality agreement)
Customer acceptance criteria: UAW0377/00
9. Philips’ Flat Panel Monitors Pixel Defect Policy
Philips’ Flat Panel Monitors Pixel Defect Policy
BRIGHT DOT DEFECTS
ACCEPTABLE LEVEL
1 lit sub-pixel
3
2 adjacent lit sub-pixels
1
3 adjacent lit sub-pixels (one white pixel)
0
Distance between two bright dots
15mm
Bright dot defects within 20 mm circle
0
Total bright dot defects of all type
3
BLACK DOT DEFECTS
ACCEPTABLE LEVEL
1 dark sub-pixel
5
2 adjacent dark sub-pixels
2
3 adjacent dark sub-pixels (one white pixel)
1
Distance between two black dots
15mm