PHILIPS Respironics Dreamstation User Manual Download Page 11

9

 Note:

 Pay close attention to all corners and crevices.

 Note: A brush is not necessary for the inside of the tube of the heated tubing. 

2. Rinse the items separately by immersing each in at least 3 gallons (11.4  liters) of water and agitate vigorously.

  Note: The tank, seal and heated tubing should each use a fresh 3 gallons (11.4  liters) of water and cannot be 

rinsed together.

3. Remove each item from the water and allow the water to drain from tank, seal and heated tubing.   

4. Air dry out of direct sunlight. 

5. Visually inspect the tank, seal and heated tubing for cleanliness. Repeat the cleaning if not visually clean.

Disinfection

The recommended disinfection methods are identified below, and can be used for a maximum of 60 cycles.
  • Thermal Disinfection: Immersion in a (tap) water bath at 75° C ± 2° C for 30 minutes
  • Cidex
  • Cidex OPA

 Caution:

 Follow all instructions from the manufacturer of the treatment products. Any deviation from these 

instructions, the manufacturer’s instructions, or agents not listed in this guide may impact the performance of the 
product. Review all applicable instructions for additional warnings and cautions.

 Caution: 

Treat the tank, seal and heated tubing as different medical devices when rinsing them with water. For 

example, the tank, seal and heated tubing cannot be rinsed in the same volume of water.

 Caution: 

Use 3 gallons (11.4 liters) as the rinse volume when following the instructions.

After Disinfection

1. Inspect the tank, seal and heated tubing for damage or wear (cracking, crazing, tears, or damage, etc.). Discard and 

replace if damaged.

2. After the final water rinse described in the chemical disinfection instructions, separately rinse the tank, seal and 

heated tubing a minimum of 1 time each in at least 3 gallons (11.4 liters) of water and air dry out of direct sunlight. 

 Note: Discoloration of the seal is normal after disinfection.

Service

The humidifier does not require routine servicing. If any part of the humidifier is worn or damaged, contact Philips 
Respironics or your home care provider. See the “Troubleshooting” section in this manual for additional information.

Traveling with the System

Packing the System

1. Remove the water tank and empty all water and allow to air dry.
2. Put the empty water tank back into the device.
3. Pack your humidifier in your carry-on luggage. 

 Caution:

 Do not move the humidifier while the water tank has water in it.

When you are traveling, the optional carrying case for your therapy device can be used for carry-on luggage only. The 

carrying case will not protect the humidifier if it is put through checked baggage. 
For your convenience at airport security stations, there is a note on the bottom of the humidifier stating that it is 

medical equipment.  It may be helpful to bring this manual along with you to help security personnel understand the 
device.
If you are traveling to a country with a line voltage different than the one you are currently using with the therapy 
device, a different power cord or an international plug adapter may be required to make your power cord compatible 
with those where you visit. Contact your home care provider for additional information.

Disposal

Dispose of this device in accordance with local regulations.

Summary of Contents for Dreamstation

Page 1: ...User manual DreamStation Heated Humidifier...

Page 2: ...ting the Tubing 4 Daily Use 5 Disconnecting the Tubing 6 Disconnecting the Therapy Device 7 Checking the Humidifier Lid Seal 7 Home Cleaning Instructions Water Tank 8 Home Cleaning Instructions Humidi...

Page 3: ...and discontinue use Contact your home care provider Before cleaning the humidifier always remove from the therapy device Empty and clean the water tank daily to prevent mold and bacteria growth Repair...

Page 4: ...ormance tubing 15 mm non heated performance tubing or optional 12 mm performance tubing For information related to the mask to be used including any need for a separate exhalation device consult the i...

Page 5: ...nal The optional heated tube connects from the humidifier to the patient s mask 10 Humidifier Connector End Connect this end of the tubing to the humidifier 11 Heater Plate Warms the water in the wate...

Page 6: ...against each other Connecting the Tubing 1 To attach the heated tube to the heated humidifier line up the connector 1 at the top of the heated tube to the top of the air outlet port 2 on the humidifie...

Page 7: ...k with water 2 no higher than the maximum fill line To avoid spilling the water level should not rise above the bottom of the tube 3 in the back of the water tank Note Clean the water tank before firs...

Page 8: ...having trouble with your mask refer to the instructions included with the mask 8 Refer to your therapy device manual for complete instructions on how to adjust both the Heated Humidifier and the opti...

Page 9: ...ompletely on both sides Allow the seal to air dry Inspect the seal for damage If the humidifier lid seal show signs of wear or damage contact your home care provider for a replacement To install or re...

Page 10: ...er plate by wiping it with a damp cloth Allow the platform to air dry before reconnecting to the therapy device 2 Inspect the humidifier base for any damage and replace it if necessary 3 Clean the hum...

Page 11: ...ng tears or damage etc Discard and replace if damaged 2 After the final water rinse described in the chemical disinfection instructions separately rinse the tank seal and heated tubing a minimum of 1...

Page 12: ...and humidifier lid seal are in present in their proper orientation and fully seated If a seal has come loose gently press around the edges of the seal to reseat it The humidifier is cracked or damage...

Page 13: ...ier seals and tube are connected properly and not leaking You may also choose to lower your humidifier and or heated tube settings or change the humidification mode from Fixed to Adaptive humidificati...

Page 14: ...eated humidifier is used with a Philips Respironics therapy device AC Power Consumption with 80W power supply 100 240 VAC 50 60 Hz 2 0 1 0 A DC Power Consumption 12 VDC 6 67 A Type of Protection Again...

Page 15: ...ized distributors of Respironics Inc products and Respironics Inc reserves the right to charge dealers for warranty service of failed product not purchased directly from Respironics Inc or authorized...

Page 16: ...1120833 1120833 R04 HM 12 05 2018 EN DOM...

Reviews: