19
SETTING THE CHAMBER TEMPERATURE
Set the chamber temperature according to the condition of use. This refrigerator can keep the stored
items for long period under appropriate temperature.
Setting range of chamber temperature: between 2
o
C and 23
o
C
Initial setting (factory setting): 5
o
C
<Important>
The chamber temperature lower than 3
o
C may cause partial freeze of stored items.
Example: Change the chamber temperature to 4
o
C from 5
o
C
► Following shows a sample setting. Set the desired chamber temperature according to the condition of
use.
Operation
Key operated
Display after operation
1
Connect the power supply plug to
the dedicated outlet.
(only when start-up of unit)
-----
The current chamber temperature
blinks.
2 Press the “set” key.
SET
The current setting (005) is displayed
and the first digit blinks.
3
Press the up arrow key and scroll
the figure to 4.
The display is changed to 004 from
005.
4 Press the “set” key.
SET
Set temperature is memorized and
the current chamber temperature
blinks.
The unit will return from setting mode to temperature display mode automatically after 90 seconds if no
key is operated (auto-return function). In this case, the setting is not changed.
<Attention>
It is recommended to install a temperature recorder (optional component) to check the highest/lowest
chamber temperature when the item which needs severe temperature control is stored.