32
VARIOUS SETTING (MENU/Tools)
Set of date, time, log interval (Tools/Date Time)
Choose the “Date Time” on the Select Tools screen to display the Date Time screen. Then set each
parameter.
Each parameter setting range:
1. Date:
Present date is indicated. (Example : in case of December 1, 2010) (Refer to a table 10.)
2. Time:
Present time is indicated. (Example : in case of 12:00:00) (Refer to a table 10.)
3. Door Delay:
The variable range of the door alarm delay is 0 minute to15 minutes. (Factory setting is
2 minutes.) (Page 34 refer to a table 11)
4.
Log Interval:
Settable between 2 minutes and 30 minutes. (Factory setting is 15 minutes.) With 15
minutes interval, recording for about 5 weeks is available.
5. Comp Delay:
Delay time for high temperature side/low temperature side compressor after power
failure. The settable range is between 3 minutes until 15 minutes. (Factory setting is 3 minutes.)