28
1
2
VARIOUS SETTING (MENU/Log)
Display of log (Log)
This product has the function to record chamber temperature. (Initialization value is for about 5 weeks by
the record interval for fifteen minutes.) This record is seen by the log indication (Log) of the display, or
data can be transmitted to the personal computer. A setup example is shown in the following.
Table 8. Display indication and the process to transmit past 1-day chamber temperature record
value
Description
of
operation
Key
operated
Indication after operation
The indication of the log data
1
-----
Top screen is displayed.
2
Press the MENU button.
MENU
The menu window is indicated. And the “set” is
highlighted.
3
Select Log by pressing
shift
key (downward)
, and press
the ENTER key.
ENTER
The record value of 1-day of chamber temperature is
indicated by the graph in the display.
4
The shift key is pressed, and
the graph of a date to transmit
to the PC is indicated.
The date of the graph changes, and a graph indicates
log data on the kept past in every 1-day.
The tra
nsmission
of the log data
5
Press the MENU button.
MENU
A menu window is indicated, and the log of 1-day (PC
1D)
is highlighted.
(Refer to the figure below.)
6
Press the Enter key when the
“PC 1D” in the menu window is
highlighted.
ENTER
The screen (Progress) which transmits log data on
chamber temperature of 1-day doing a record to the
personal computer is indicated.
7
Highlight the “Start” in the menu
window and press the ENTER
key.
ENTER
Transmission is started in the personal computer of the log
data on chamber temperature of 1-day being recorded.
"Sending" and the number of transmit data are indicated in
the display at the time of the transmission.
8
A completion message (Finished.) is indicated.
The end of
the lo
g data display
9
If a completion message is
indicated, press the MENU
button.
MENU
The menu window is indicated. And the “Start” is
highlighted.
10
The
shift key (downward)
is
pressed, and Cancel of the
menu window is chosen, and
press the ENTER key.
ENTER
Log data on 1-day of chamber temperature is indicated
by the graph in the display.
11 Press the MENU button.
MENU
The record (PC 1D) of 1-day of the menu window is
highlighted.
12
Press the shift key (downward)
to highlight the “Cancel” in the
menu window, and press the
ENTER key.
ENTER
Top screen is displayed.
Note:
When this function is used, an optional interface board (MTR-480 only) is necessary. Contact our
sales representative or agent at the time of installation of the optional parts.
1. The indication of the date of the log screen:
The log date of the chamber temperature data is
indicated. When a log data screen is indicated and the left-right key of the shift key is pressed, the date
of 1. changes, and the chamber temperature log of that date is indicated. Indicate the log screen of a
date to transmit when data on 1-day are transmitted to the PC, and start transmission.
2. Temperature indication range:
When the up-down key of the shift key is pressed, indication
temperature changes from -100
o
C - -25
o
C to -25
o
C - 50
o
C.