background image

 

 

If the coupling status indicator does not appear not stable, or the numbers on the display seem erratic, check to 

make  sure  that  there  is  an  adequate  film  of  couplant  beneath  the  transducer,  and  that  the  transducer  is  seated  flat 
against the material. If the condition persists, it may be necessary to select a different transducer (size or frequency) 
for the material being measured. 

While  the  transducer  is  in  contact  with  the  material  that  is  being  measured,  the  instrument  will  perform  four 

measurements every second, updating its display as it does so. When the transducer is removed from the surface, the 
display will hold the last measurement made. 

Note

Occasionally,  a  small  film  of  couplant  will  be  drawn  out  between  the  transducer  and  the  surface  as  the 

transducer  is  removed.  When  this  happens,  the  gauge  may  perform  a  measurement  through  this  couplant  film, 
resulting  in  a  measurement  that  is  larger  or  smaller  than  it  should  be.  This  phenomenon  is  obvious  when  one 
thickness  value  is  observed  while  the  transducer  is  in  place,  and  another  value  is  observed  after  the  transducer  is 
removed.  In  addition,  measurements  through  very  thick  paint  or  coatings  may  result  in  the  paint  or  coating  being 
measured  rather  than  the  actual  material  intended.  The  responsibility  for  proper  use  of  the  instrument,  and 
recognition of these types of phenomenon, rests solely with the user of the instrument. 

4.5

 

Scan

 

mode

 

While the gauge excels at making single point measurements, it is sometimes desirable to examine a larger region, 

searching for the thinnest point. The gauge includes a feature, called Scan Mode, which allows it to do just that.   

In normal operation, the gauge performs and displays four measurements every second, which is quite adequate 

for single measurements. In Scan Mode, however, the gauge performs ten measurements every second, and displays the 
readings while scanning. While the transducer is in contact with the material being measured, the gauge is keeping 
track  of  the  lowest  measurement  it  finds.  The  transducer  may  be  “scrubbed”  across  a  surface,  and  any  brief

 

interruptions  in  the  signal  will  be  ignored.  When  the  transducer  loses  contact  with  the  surface  for  more  than  two 
seconds, the gauge will display the smallest measurement it found. When the transducer is removed from the material 
being scanned, the gauge will display the smallest measurement it found. 

When the scan mode is turned off, the single point mode will be automatically turned on. Turn on/off the scan 

mode by the following steps: 

 
Press the 

  key to switch the scan measurement mode on and off. It will display the current condition of the 

scan mode on the main screen. 

 

4.6

 

Changing

 

Resolution

 

UTG‐2900 has selectable display resolution, which is 0.01” / 0.001” (0.1mm and 0.01mm).   
Press the 

  key    and the gauge will switch the resolution between “High” and “Low”. 

4.7

 

Changing

 

Measuring

 

Units

 

On the measurement mode, press the 

  key to switch back and forth between inch and metric units.   

 
 
 
 
 
 
 
 
 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for UTG-2900

Page 1: ...Model No UTG 2900 283 Veterans Blvd Carlstadt NJ 07072 201 933 6300 www phase2plus com www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 2: ...9 4 2 Probe Zero 9 4 3 Sound Velocity Calibration 10 4 4 Making Measurements 12 4 5 Scan mode 13 4 6 Changing Resolution 13 4 7 Changing Units 13 4 8 Memory Management 14 4 9 Data Printing 15 4 10 Bee...

Page 3: ...ring Range Pulse Echo 0 025 23 0 in steel Echo Echo 0 118 2 35 Accuracy 0 5 thickness 001 Sound velocity range 3280 32805 ft s 1000 9999m s Upper Lower Limit Preset Alarm Yes Scanning Feature Yes 16 m...

Page 4: ...ickness of a part or structure by accurately measuring the time required for a short ultrasonic pulse generated by a transducer to travel through the thickness of the material reflect from the back or...

Page 5: ...Indicates the coupling status While the gauge is taking a measurement the coupling indicator will be displayed If it is not the gauge is having difficulty achieving a stable measurement and the thick...

Page 6: ...2 Save 3 Enter 4 Back Light 5 Probe Zero 6 Inch metric 7 Thru Coating on off 8 Scan on off 9 Calibration Velocity setting 10 Calibration block 0 157 4mm 1 5 6 7 3 2 8 9 4 10 www GlobalTestSupply com F...

Page 7: ...f the material through which the sound travels has any grain structure the sound waves will experience scattering Both of these effects reduce the strength of the waves and thus the gauge s ability to...

Page 8: ...ing measured and the other semicircle is responsible for conducting the echoed sound back into the transducer When the transducer is placed against the material being measured it is the area directly...

Page 9: ...Transducers should be inspected on a regular basis for signs of uneven wear of the wearface If the wearface is worn on one side more than another the sound beam penetrating the test material may no l...

Page 10: ...r factor and will compensate for this value in any subsequent measurements When performing a probe zero the instrument will always use the sound velocity value of the built in probe disc even if some...

Page 11: ...on mode and return to the measurement mode The gauge is now ready to perform measurements 4 3 2 Calibration to a known velocity Note This procedure requires that the operator knows the sound velocity...

Page 12: ...locities to enable the user to quickly check different known materials Example If you are testing steel at 233 in us you can quickly change to one of your stored sound velocities when you need to test...

Page 13: ...gauge excels at making single point measurements it is sometimes desirable to examine a larger region searching for the thinnest point The gauge includes a feature called Scan Mode which allows it to...

Page 14: ...the key to activate the data logging functions It will display the current file name and the total record count of the file 2 Use the key and the key to scroll to the file that will be cleared of all...

Page 15: ...background light at any moment as you need after power on Since the EL light will consume more power turn on it only when necessary 4 12 Battery Icon Two AA size alkaline batteries are used as the pow...

Page 16: ...f repairs must be supplied as well ANY ATTEMPT AT HOME REPAIR WILL AUTOMATICALLY VOID THE STATED WARRANTY NO EXCEPTIONS 6 Transport and Storage Be sure to clean the probe and cable after each use Grea...

Page 17: ...ommended that the probe be left in contact with the surface for as short a time as needed to acquire a stable measurement While the transducer is in contact with a hot surface it will begin to heat up...

Page 18: ...thus may be undesirable Other suitable couplants for measurements at normal temperatures may include water various oils and greases gels and silicone fluids Measurements at elevated temperatures will...

Page 19: ...85 2200 Magnesium 0 230 5800 Mercury 0 057 1400 Molybdenum 0 250 6300 Monel 0 210 5400 Neoprene 0 063 1600 Nickel 0 220 5600 Nylon 6 6 0 100 2600 Oil SAE 30 0 067 1700 Platinum 0 130 3300 Plexiglass 0...

Reviews: