Symptom / malfunction messages
Possible causes
Remedy
Filament potential error
Filament voltage too high (filament
open)
Filament voltage too low (filament
shorted)
(Filament potential error)
Insufficient vacuum
● Ensure that the pressure is below the maximum
operating pressure: < 5 × 10
-4
hPa.
Analyzer filament defec-
tive (broken or short-
circuited)
● Check the ion source elements and analyzer ele-
ments for short-circuit and insulation.
● Ensure that the wiring screws are securely fas-
tened.
● Ensure that the flange or the pipe features an in-
ner diameter of ≥ 37 mm, preferably > 39 mm.
● Replace the filament unit, the ion source or the
analyzer.
Poor contact of wiring
● Check the ion source elements and analyzer ele-
ments for short-circuit and insulation.
● Ensure that the wiring screws are securely fas-
tened.
● Ensure that the flange or the pipe features an in-
ner diameter of ≥ 37 mm, preferably > 39 mm.
● Replace the filament unit, the ion source or the
analyzer.
Filament current error
Filament current too high (filament
shorted)
Filament current too low (filament
open)
(Filament current error)
Insufficient vacuum
● Ensure that the pressure is below the maximum
operating pressure: < 5 × 10
-4
hPa.
Analyzer filament defec-
tive (broken or short-
circuited)
● Check the ion source elements and analyzer ele-
ments for short-circuit and insulation.
● Ensure that the wiring screws are securely fas-
tened.
● Ensure that the flange or the pipe features an in-
ner diameter of ≥ 37 mm, preferably > 39 mm.
● Replace the filament unit, the ion source or the
analyzer.
Poor contact of wiring
● Check the ion source elements and analyzer ele-
ments for short-circuit and insulation.
● Ensure that the wiring screws are securely fas-
tened.
● Ensure that the flange or the pipe features an in-
ner diameter of ≥ 37 mm, preferably > 39 mm.
● Replace the filament unit, the ion source or the
analyzer.
Focus error
It is not possible to start/maintain the
focus voltage.
(Focus error)
Focus plate shorted
● Check the ion source elements and analyzer ele-
ments for short-circuit and insulation.
● Ensure that the wiring screws are securely fas-
tened.
● Ensure that the flange or the pipe features an in-
ner diameter of ≥ 37 mm, preferably > 39 mm.
● Replace the IS 250.
Focus voltage out of ex-
pected range
● Check the ion source elements and analyzer ele-
ments for short-circuit and insulation.
● Ensure that the wiring screws are securely fas-
tened.
● Ensure that the flange or the pipe features an in-
ner diameter of ≥ 37 mm, preferably > 39 mm.
● Replace the IS 250.
Internal power supply failure
Error in the electronic unit
(QME)
● Contact
Tbl. 16:
Hardware errors
Malfunctions