26
BG 805 204 BE / B (0203) QMS200.oi
When the Prisma is used for the first time, the factory parameter settings of the ion
source are applied. TalkStar reads these parameter values when it starts
communication with the Prisma.
!
Start TalkStar. In the tree, select "Recipes" and start "Tune Ionsource" with
a double click.
"
In the context menu of "Devices", click "Properties" with the right mouse
button. In the window that is subsequently displayed, choose the
"Ionsource" page.
5.2.5 Optimizing the Ion
Source Sensitivity
A scan in the 26 … 30 amu mass
range is displayed. This mass scan
can be used for optimizing the sensi-
tivity and mass separation of the ion
source.