22
BG 805 204 BE / B (0203) QMS200.oi
5 Maintenance
Skilled personnel
Filaments are supplied as preassembled units. They should only be changed by
qualified personnel.
Caution
Caution: dirt sensitive area
Touching the product or parts thereof with one's bare hands increases
the desorption rate.
Always wear clean, lint-free gloves and use clean tools when working
in this area.
Mounting tool
!
Detach all electrical connec-
tions to the QME.
"
Separate the QME from the
analyzer (by performing the
steps in Chapter 3.2.4 in re-
verse order).
#
Detach the analyzer from the
vacuum chamber.
$
Place the analyzer into the
supplied mounting tool.
%
Grip the U-section with flat-
nose pliers and loosen the
screw by 2 turns.
&
Remove the nut.
'
Carefully pull the filament
assembly upward.
(
Dispose of the used filament
assembly.
5.1 QMA
5.1.1 Changing the Filament