![Pfeiffer Vacuum Prisma 80 QMS 200 Operating Instructions Manual Download Page 16](http://html1.mh-extra.com/html/pfeiffer-vacuum/prisma-80-qms-200/prisma-80-qms-200_operating-instructions-manual_1534984016.webp)
16
BG 805 204 BE / B (0203) QMS200.oi
4 Operation
DANGER
Caution: damaged product
Putting a damaged product into operation can be extremely hazard-
ous.
In case of visible damages, make sure the product is not put into
operation.
DANGER
Caution: line voltage
Incorrectly grounded products can be extremely hazardous in the
event of a fault.
Use only a 3-conductor power cable with protective ground. The power
connector may only be plugged into a socket with a protective ground.
The protection must not be nullified by an extension cable without
protective ground.
WARNING
Check the pressure in the QMA before you insert the power plug of the
SP 200. The pressure must be <10
-4
mbar.
WARNING
Do always make sure that the limits values specified in the Technical
Data (
→
"
6) are not exceeded and that the ambient conditions are
met.
Insert the power plug.
WARNING
The power plug should be easily accessible. If this is not the case at the instal-
lation site, a power switch or connector must be provided at an easily accessible
location.
4.1 First Time Operation
SP 200