Operating fluid type
Fundamentally, when filling, refilling, or changing operating fluid, you must always use the
operating fluid type specified on the rating plate. Should process conditions change, you
can convert to a different operating fluid type.
Safety data sheets
You can obtain the safety data sheets for operating fluids from Pfeiffer Vacuum on request,
or from the
Pfeiffer Vacuum Download Center
The usable life of operating fluid is dependent on the area of application of the rotary vane vacuum
pumps.
Instructions for when operating fluid should be changed
● Vacuum pump does not reach the specified final pressure.
● Operating fluid soiled, milky or unclear
● Mineral operating fluid with thermic aging (color number > 4)
7.4.1 Determine the degree of aging of P3 operating fluid
WARNING
Health hazard and risk of environmental damage from toxic contaminated operating fluid
Toxic process media can cause operating fluid contamination. When changing the operating fluid,
there is a health hazard due to contact with poisonous substances. Illegal disposal of toxic substan-
ces causes environmental damage.
► Wear suitable personal protective equipment when handling these media.
► Dispose of the operating fluid according to locally applicable regulations.
You can determine the degree of aging of P3 operating fluid in clean processes using the color chart (in
accordance with DIN 51578). The supplementary sheet with the document number PK0219 can be ob-
tained from the
Pfeiffer Vacuum Download Center
Prerequisites
● Vacuum pump switched off
● Vacuum pump is vented to atmospheric pressure on the suction side
● Vacuum pump cooled
Required aids
● Test tube
● Pipette with flexible hose
● Allen key,
WAF 8
● Calibrated torque wrench (tightening factor ≤ 2.5)
Determine degree of aging of P3 operating fluid
1. Unscrew the filler screw.
– Be careful with the O-ring.
2. Use the pipette to extract a sample of the operating fluid from the filling port.
3. Pour the sample into a test tube.
4. Check the sample in bright light.
5. If it is a reddish-brown color at the latest (corresponding with color identification number 5),
change the operating fluid.
6. Screw in the filler screw.
– Be careful with the O-ring.
– Tightening torque:
6 Nm
Maintenance
36/60