Condensate separator
Pfeiffer Vacuum recommends installing a condensate separator, with condensate drain at
the lowest point of the exhaust line.
Installation and operation of accessories
Pfeiffer Vacuum offers a series of special, compatible accessories for its rotary vane
pumps.
● You can find information and ordering options for approved
● Described accessories are not included in the shipment.
1
3
4
5
2
Fig. 7:
Exhaust connection with flange connection
1 Exhaust line
4
O-ring
2 Protective cap
5
Circlip
3 Centering ring with cone sieve
Procedure
1. Remove the protective cap from the exhaust flange.
2. Ensure that the centering ring with cone sieve is fitted in the exhaust flange.
3. Choose a minimum exhaust line diameter equal to the nominal diameter of the connection flange.
4. Depending on the pump type, use PVC or metallic hoses with flange connections from the
5. Route the piping downwards from the vacuum pump, to prevent condensate return.
6. Support or suspend the piping to the vacuum pump so that no piping system forces act on the
vacuum pump.
7. Connect both flanges with a circlip.
5.4 Establishing mains connection
DANGER
Danger to life from electric shock
Touching exposed and voltage-bearing elements causes an electric shock. Improper connection of
the mains supply leads to the risk of touchable live housing parts. There is a risk to life.
► Before the installation, check that the connection leads are voltage-free.
► Make sure that electrical installations are only carried out by qualified electricians.
► Provide adequate grounding for the device.
► After connection work, carry out an earthed conductor check.
Installation