Risks during installation
DANGER
Danger to life from electric shock
Touching exposed and voltage-bearing elements causes an electric shock. Improper connection of
the mains supply leads to the risk of touchable live housing parts. There is a risk to life.
► Before the installation, check that the connection leads are voltage-free.
► Make sure that electrical installations are only carried out by qualified electricians.
► Provide adequate grounding for the device.
► After connection work, carry out an earthed conductor check.
WARNING
Risk of crushing from rotating parts
Fingers and hands may be caught by rotating pistons within the connection flange. This results in se-
vere injuries.
► Keep limbs out of the reach of the roots pump.
CAUTION
Danger of injury from bursting as a result of high pressure in the exhaust line
Faulty or inadequate exhaust pipes lead to dangerous situations, e.g. increased exhaust pressure.
There is a danger of bursting. Injuries caused by flying fragments, the escaping of high pressure, and
damage to the unit cannot be excluded.
► Route the exhaust line without shut-off units.
► Observe the permissible pressures and pressure differentials for the product.
► Check the function of the exhaust line on a regular basis.
CAUTION
Risk of injury from loss of stability
During setup, there is a risk of injury from tipping, if the vacuum pump is not anchored on the stand-
ing surface.
► Secure the vacuum pump using suitable lifting gear.
► Wear personal protective equipment.
CAUTION
Danger of injury from moving parts
After a power failure or a standstill as a result of overheating, the motor restarts automatically. There
is a risk of injury to fingers and hands if they enter the operating range of rotating parts.
► Safely disconnect motor from the mains.
► Secure the motor against reactivation.
► Dismantle the vacuum pump for inspection, away from the system if necessary.
Risks during operation
WARNING
Danger of poisoning due to toxic process media escaping from the exhaust pipe
During operation with no exhaust line, the vacuum pump allows exhaust gases and vapors to escape
freely into the air. There is a risk of injury and fatality due to poisoning in processes with toxic process
media.
► Observe the pertinent regulations for handling toxic process media.
► Safely purge toxic process media via an exhaust line.
► Use appropriate filter equipment to separate toxic process media.
Safety
11/52