7 Maintenance
7.1 General maintenance information
WARNING
Danger to life from electric shock during maintenance and service work
The device is only completely de-energized when the mains plug has been disconnected and the tur-
bopump is at a standstill. There is a danger to life from electric shock when making contact with live
components.
► Before performing all work, switch off the main switch.
► Wait until the turbopump comes to a standstill (rotation speed =0).
► Remove the mains plug from the device.
► Secure the device against unintentional restarting.
WARNING
Health hazard through poisoning from toxic contaminated components or devices
Toxic process media result in contamination of devices or parts of them. During maintenance work,
there is a risk to health from contact with these poisonous substances. Illegal disposal of toxic sub-
stances causes environmental damage.
► Take suitable safety precautions and prevent health hazards or environmental pollution by toxic
process media.
► Decontaminate affected parts before carrying out maintenance work.
► Wear protective equipment.
WARNING
Risk of cuts on moving, sharp-edged parts when reaching into the open high vacuum connec-
tion
Incorrect handling of the turbopump before maintenance work results in hazardous situations with
risk of injury. There is a risk of cuts from accessing sharp-edged, rotating parts when removing the
turbopump.
► Wait until the turbopump comes to a standstill (rotation speed f=0).
► Switch the turbopump off properly.
► Secure the turbopump against re-start.
► Seal open connections immediately following removal, using the original protective cover.
7.2 Checklist for inspection and maintenance
Notes on maintenance intervals
Depending on the process, the required maintenance intervals may be shorter than the ref-
erence values specified in the table.
● Consult with Pfeiffer Vacuum Service about shorter maintenance intervals for extreme
loads or for specific processes.
Maintenance Level 2 and Level 3
We recommend Pfeiffer Vacuum Service (PV) carry out maintenance work at Level 2 and
Level 3 (inspection). If the specified intervals are exceeded, or if maintenance work is car-
ried out improperly, no warranty or liability claims are accepted on the part of Pfeiffer Vac-
uum. This also applies wherever parts other than original spare parts are used.
Recommendations for performing maintenance measures
► You can carry out maintenance work at
Level 1
yourself.
► Utilize a lint-free cloth and a little isopropanol for cleaning.
Maintenance
37/58