Infringement of conformity due to modifications to the product
The Declaration of Conformity from the manufacturer is no longer valid if the operator
changes the original product or installs additional equipment.
● Following the installation into a system, the operator is required to check and re-evalu-
ate the conformity of the overall system in the context of the relevant European Direc-
tives, before commissioning that system.
General safety precautions
► Do not expose body parts to the vacuum.
► Observe the safety and accident prevention regulations, if necessary wear personal protective
equipment.
► Check all safety measures at regular intervals.
► Always ensure a secure connection to the earthed conductor (PE), protection class I.
► During operation, make sure that plug-and-socket connections are securely fitted.
► Never operate the vacuum pump with open vacuum flange.
► Never make your own conversions or modifications to the vacuum pump.
► Before returning the vacuum pump, observe the notes in the chapter Service.
2.4 Limits of use
Parameter
HiLobe
Installation location
weatherproof (internal space)
Installation altitude
max. 2000 m above sea level
1)
Orientation
horizontal, max. permissible angle of inclina-
tion: ± 3 °
Ambient temperature
+5
℃
to +40
℃
Relative humidity of air
max. 85 %
Permanent intake pressure in circulation mode < 1100 hPa (abs.)
Permanent intake pressure in vacuum mode
depending on max. pressure differential
Max. gas temperature, pressure side
250
℃
Pumping speed ratio to backing pump
≤ 10:1
Degree of pollution
2
Protection class
I
Overvoltage category
II
Permissible protection degree
IP54
Type 12 according to UL 50E
Tbl. 3:
Permissible ambient conditions
1)
at installation altitudes > 1,000m above m.s.l. and an ambient temperature of 40 °C the rated power of the
motor reduces by around 10 %
Safety
16/68