Procedure
1. Remove the protective cap from the exhaust connection.
2. With vertical installation, connect the exhaust line directly with the exhaust connection.
3. With horizontal installation (180°), connect the exhaust line using an outlet elbow.
4. Route the piping downwards from the vacuum pump, to prevent condensate return.
5. Install a condensate separator if necessary.
5.4 Connecting the cooling water supply
Optional monitoring systems
● Flow rate monitor
● Cooling water pressure monitor
Prerequisites
● Cooling water connection unpressurised
● Cooling system evacuated
1
2
Fig. 6:
Cooling water connection
1 Cooling water inlet
2
Cooling water outlet
Connecting the cooling water supply
1. Ensure that the cooling water flow is monitored at the cooling water outlet.
– If this is not possible, calibrate the water pressure with one or two orifice plates.
2. Connect the cooling water connections with the water supply.
3. Open the water inflow.
Parameter
Cooling water
Appearance
● filtered
● mechanically clear
● visually clear
● no turbidity
● no sediment
● free from grease and oil
pH value
7 to 8
Carbonate hardness, max.
10 °dH
12.53 °e
17.8 °fH
178 ppm CaCO
3
Chloride content, max.
100 mg/l
Sulfate content, max.
240 mg/l
Carbonic acid content, max.
not detectable
Ammonia content, max.
not detectable
Installation
22/52