background image

7

4.5. Gas Ballast Valve (For Corrosive Gas

Processes)

Gas ballast valve 7 is equipped with a dosing unit 7.3 and a
flushing gas connection 7.4.
The dosing unit is closed when turned fully clockwise and
open when turned fully anti-clockwise. The amount of gas
can be set at any required rate from maximum to minimum
with a flow meter.

Gas ballast valve 7, 

Standard version

Position: Closed

Position: Open

Please note:

Flushing gas must be connected when operating corrosive
gas model pumps. The gas ballast valve must be open.
Flushing gas pressure may not exceed 1.5 bar. The type of
flushing gas is dependent on the medium being pumped. Our
specialists would be pleased to give advice regarding the
type and volume of flushing gas required.

When working with synthetic pump fluid, the
manufacturer´s directions for use must be
observed. If F5 is heated to over 250°C,
poisonous gases damaging to the respiratory
system are produced. F5 must not be allowed to

come into contact with tobacco products. When handling
chemicals, the relevant manufacturer´s instructions must be
followed. Pump fluid which have been enriched with toxic or
corrosive substances must be specially handled and all perti-
nent regulations must be observed.

4.6. Silencer

Silencer 40 is a nozzle inside the pump housing and cannot be
altered; when dirty it should either be cleaned or replaced
(see Service Instructions).

7.1

7.2

7.4

Gas ballast valve 7, for Corrosive Gas Processes

1

Vacuum connection

2

Exhaust connection

3

Operating fluid filler screw

4

Operating fluid drain screw

5

Sight glass

6

Motor

7

Gas ballast valve

8

Support stand

9

Casing

44 Base plate

Working principle of High Vacuum Safety Valve 20

4.4. Gas Ballast Valve (Standard Version)

To avoid the condensation of vapours in the pump, air is peri-
odically fed into the working chamber at the beginning of the
compression phase via the gas ballast valve.

– Gas ballast valve 7 is shut when turning to the right to posi-

tion 0–0 and open when turning to the left to position 1–1.
Intermediate settings are not possible.

If the pumping process requires the connection of flushing
gas, an adapter connection G 1/8 can be provided for the gas
ballast valve (available on request).

21

23

26

Air

24

27

22

25

Oil

53

7.3

7.1 O-ring
7.2 O-ring
7.3 dosing unit
7.4 Flushing gas 

connection
(for 9 mm hose)

To avoid condensation in the pump, vapours
should only be pumped off when the pump is
warm and with gas ballast valve open. Note
pump water vapour compatibility levels under
Section 8. Technical Data. When the process

has been completed, allow the pump to continue running for
about 30 minutes with open gas ballast valve for operating
fluid regeneration purposes.

Summary of Contents for DUO 2.5 A

Page 1: ...Operating Instructions Rotary Vane Vacuum Pumps UNO 005 A UNO DUO 2 5 A UNO DUO 2 5 AC PK 800 152 BGB F 9911...

Page 2: ...em Modifications reserved Danger of burns from touching hot parts 1 Safety Precautions Danger of personal injury 1 Safety Precautions 2 1 1 For Your Orientation 2 2 Understanding The Pump 3 2 1 Main F...

Page 3: ...um connection and the pumping system The valve closes off the vacuum equipment without delay when the pump is at a standstill and vents the pump The Rotary Vane Vacuum Pumps are in two versions availa...

Page 4: ...uctance value of the accessory there is a reduction in volume flow rate For the intake lines metal or PVC hoses with small flanges are available for laboratory work 3 4 Connecting The Exhaust Side Pre...

Page 5: ...values The single phase drive motors of the UNO 005 A 2 5 A and the DUO 2 5 A AC are fitted with a built in thermal protection switch A cable with connecting plug is fitted to the motor for mains con...

Page 6: ...id is possible when the pump is in final vacuum operations If there is a noticeable loss of pump fluid the casing and the radial shaft seals should be checked for leaks see Service Instructions 4 2 Sw...

Page 7: ...cer 40 is a nozzle inside the pump housing and cannot be altered when dirty it should either be cleaned or replaced see Service Instructions 7 1 7 2 7 4 Gas ballast valve 7 for Corrosive Gas Processes...

Page 8: ...P3 can be read off from the colour scale in accordance with DIN 51578 on the supplementary sheet PK 800 219 BN B on request The colour scale enables precise colour determination Fill a specimen in a...

Page 9: ...erials in accordance with the Code of Federal Regulations CFR 49 Part 173 240 Definition and Preparation We will carry out the decontamination and invoice this work to you if you have not attached thi...

Page 10: ...mbar 5 10 2 5 10 2 5 0 10 3 Total with gas ballast mbar 1 1 6 0 10 3 Max pressure in continuous operation bar 1 5 1 5 1 5 Water vapour compatibility mbar 20 15 15 Water vapour capacity at 50 60 Hz g...

Page 11: ...s catalogue contains details of a comprehensive accessory programme which can both extend the range of application of the pump and provide protection against wear and tear under extreme operating cond...

Page 12: ...full part number When ordering spare parts please state additionally the unit type and unit number see rating plate Please use this list as an order form by taking a copy Spare parts UNO 005 A Standar...

Page 13: ...he repair and or service of vacuum components will only be carried out if a correctly completed declaration has been submitted Non completion will result in delay The manufacturer could refuse to acce...

Page 14: ...tate in basso le direttive aplicate le norme standardizzate nonch le norme e le specifiche nazionali utilizatte E E E ES S S S Declaraci n de conformidad al tenor de las siguientes Directivas de la UE...

Page 15: ...G G GR R R R Produkt Product Produit Prodotto Producto Produkt Produkt Produto UNO 005 A UNO DUO 2 5 A UNO DUO 2 5 AC Angewendete Richtlinien harmonisierte Normen und angewendete nationale Normen in...

Page 16: ...2 259 6243 U S A Pfeiffer Vacuum Inc 24 Trafalgar Square Nashua NH 03063 1988 USA telephone 1 603 578 6500 telefax 1 603 578 6550 Venezuela Secotec S A Apartado 3452 Caracas 1010 A telephone 58 2 573...

Reviews: