The Pfeiffer Vacuum parameter set
19
Control via interface
Configuration via parameters
[P:060]
and
[P:061]
.
Digital inputs on "
X3
"
Configuration via parameters
[P:062]
and
[P:063]
.
6.4
Operation with the Pfeiffer Vacuum parameter set
Factory settings
The electronic drive unit is pre-programmed in the factory. This guarantees proper, reli-
able turbopump operation without the need for additional configuration.
Checking the adjust-
ments
Before operating with parameters, check set values and control commands for their
suitability for the pumping process.
Gas type dependent
operations
Friction causes the rotor to heat up severely under gas load and high rotation speed. To
avoid overheating, the electronic drive unit has implemented power-rotation speed-char-
acteristics, whereby the pump can be operated at every rotation speed with the maxi-
mum allowable gas load without danger of damage. The maximum power consumption
depends on the gas type. Three characteristics are available in order to completely ex-
haust the pump's capacity for each gas type.
●
Gas mode "0" for gases with the molecular mass >39, e.g. argon.
●
Gas mode "1" for gases with the molecular mass
≤
39.
●
Gas mode "2" for helium.
●
Power characteristics according to the technical data of the turbopump.
Check and set-up the gas mode via
[P:027]
.
Option [P:060]
Description
1 = remote
Operation via connection "
remote
"
2 = RS-485
Operation via connection "
RS-485
"
4 = PV.can
For service purposes only
8 = Field bus
Operation via field bus
16 = E74
Operation via connection "
E74
"
Option [P:061]
Description
0 = off
Interface selection via [P:060]
1 = on
Interface selection locked
Option
Description
0 = deaktiviert
Connection deactivated
1 = Enable venting
Control is equal to parameter [P:012]
2 = Heating
Control is equal to parameter [P:001]
3 = Sealing gas
Control is equal to parameter [P:050]
4 = Run-up time control
Control is equal to parameter [P:004]
5 = Rotation speed setting mode
Control is equal to parameter [P:026]
6 = Motor
Control is equal to parameter [P:023]
NOTICE
Danger of the pump being destroyed
Pumping of gases with a higher molecular mass in the wrong gas mode can lead to de-
struction of the pump.
Ensure the gas mode is correctly set.
Contact Pfeiffer Vacuum before using gases with a greater molecular mass (> 80).