The Pfeiffer Vacuum parameter set
15
Control commands
#
Display
Designation
Functions
Dat
a
t
yp
e
Acc
ess
Unit
min max
default
001 Heating
Heating
0 = off
1 = on
0 RW
0
1
0
x
002 Standby
Standby
0 = off
1 = on
0 RW
0
1
0
x
004 RUTimeCtrl
Run-up time control
0 = off
1 = on
0 RW
0
1
1
x
009 ErrorAckn
Error acknowledgement
1 = Error acknowledgement
0 W
1
1
010 PumpgStatn
Pumping station
0 = off
1 = on and error acknowledgement
0 RW
0
1
0
x
012 EnableVent
Enable venting
0 = no
1 = yes
0 RW
0
1
0
x
017 CfgSpdSwPt
Configuration rotation speed switchpoint
0 = Rotation speed switchpoint 1
1 = Rotation speed switchpoint 1&2
7 RW
0
1
0
x
019 Cfg DO2
Configuration output DO2
0 = Rot. speed switchpoint attained
1 = No error
2 = Error
3 = Warning
4 = Error and/or warning
5 = Set speed attained
6 = Pump on
7 = Pump accelerates
8 = Pump decelerates
9 = always 0
10 = always 1
11 = Remote priority active
12 = Heating
13 = Backing pump
14 = Sealing gas
15 = Pumping station
16 = Pump rotates
17 = Pump does not rotate
7 RW
0
17
1
x
023 MotorPump
Motor pump
0 = off
1 = on
0 RW
0
1
1
x
024 Cfg DO1
Configuration output DO1
0 = Rot. speed switchpoint attained
1 = No error
2 = Error
3 = Warning
4 = Error and/or warning
5 = Set speed attained
6 = Pump on
7 = Pump accelerates
8 = Pump decelerates
9 = always 0
10 = always 1
11 = Remote priority active
12 = Heating
13 = Backing pump
14 = Sealing gas
15 = Pumping station
16 = Pump rotates
17 = Pump does not rotate
7 RW
0
17
0
x
025 OpMode BKP
Backing pump mode
0 = Continuous operation
1 = Intermittent operation
2 = Delayed switching on
3 = Delayed intermittent operation
7 RW
0
3
0
x
026 SpdSetMode
Rotation speed setting mode
0 = off
1 = on
7 RW
0
1
0
x
027 GasMode
Gas mode
0 = Heavy gases
1 = Light gases
2 = Helium
7 RW
0
2
0
x
030 VentMode
Venting mode
0 = Delayed venting
1 = No venting
2 = Direct venting
7 RW
0
2
0
x
035 Cfg Acc A1
Configuration accessory connection A1
0 = Fan (continous operation)
1 = Venting valve, normally closed
2 = Heating
3 = Backing pump
4 = Fan (temperature controlled)
5 = Sealing gas
6 = always 0
7 = always 1
8 = Power failure venting unit
7 RW
0
8
0
x