The Pfeiffer Vacuum parameter set
Status requests
036 Cfg Acc B1
Configuration accessory connection B1
0 = Fan (continous operation)
1 = Venting valve, normally closed
2 = Heating
3 = Backing pump
4 = Fan (temperature controlled)
5 = Sealing gas
6 = always 0
7 = always 1
8 = Power failure venting unit
7 RW
0
8
1
x
037 Cfg Acc A2
Configuration accessory connection A2
0 = Fan (continous operation)
1 = Venting valve, normally closed
2 = Heating
3 = Backing pump
4 = Fan (temperature controlled)
5 = Sealing gas
6 = always 0
7 = always 1
8 = Power failure venting unit
7 RW
0
8
3
x
038 Cfg Acc B2
Configuration accessory connection B2
0 = Fan (continous operation)
1 = Venting valve, normally closed
2 = Heating
3 = Backing pump
4 = Fan (temperature controlled)
5 = Sealing gas
6 = always 0
7 = always 1
8 = Power failure venting unit
7 RW
0
8
2
x
050 SealingGas
Sealing gas
0 = off
1 = on
0 RW
0
1
0
x
055 Cfg AO1
Configuration output AO1
0 = Actual rotation speed
1 = Power
2 = Current
3 = always 0 V
4 = always 10 V
7 RW
0
4
0
x
060 CtrlViaInt
Control via interface
1 = Remote
2 = RS-485
4 = PV.can
8 = Field bus
16 = E74
255 = Unlock interface selection
7 RW
1
255
1
x
061 IntSelLckd
Interface selection locked
0 = off
1 = on
0 RW
0
1
0
x
062 Cfg DI1
Configuration input DI1
0 = Deactivated
1 = Enable venting
2 = Heating
3 = Sealing gas
4 = Run-up time control
5 = Rotation speed setting mode
6 = Motor
Setting
≠
[P:063]
7 RW
0
6
1
x
063 Cfg DI2
Configuration input DI2
0 = Deactivated
1 = Enable venting
2 = Heating
3 = Sealing gas
4 = Run-up time control
5 = Rotation speed setting mode
6 = Motor
Setting
≠
[P:062]
7 RW
0
6
2
x
#
Display
Designation
Functions
Dat
a
t
yp
e
Acc
ess
Unit
min max
default
#
Display
Designation
Functions
Dat
a
ty
pe
A
c
ce
ss
Unit
min max
default
300
RemotePrio
Remote priority
0 = no
1 = yes
0
R
0
1
302
SpdSwPtAtt
Rotation speed switchpoint attained
0 = no
1 = yes
0
R
0
1
303
Error code
Error code
4
R
304
OvTempElec
Excess temperature electronic drive unit
0 = no
1 = yes
0
R
0
1
305
OvTempPump
Excess temperature pump
0 = no
1 = yes
0
R
0
1
306
SetSpdAtt
Set rotation speed attained
0 = no
1 = yes
0
R
0
1