This could happen if the sample you try to analyze
is of very low density. If you receive this message
for a sample of normal density, please contact
technical support.
Empty weight is not stable
The empty weight measurement is not stable for
analysis. This may be caused by instrument
instability due to unstable or vibration
environment. First insure the instrument is
properly leveled on a solid surface as per section
5.1.3 of this manual. If problem persists, please
contact technical support.
Sample weight is not stable
The sample weight measurement is not stable for
analysis. This may be caused by instrument
instability due to unstable or vibration
environment. First insure the instrument is
properly leveled on a solid surface as per section
5.1.3 of this manual. If problem persists please
contact technical support.
If any of the following error messages appear, please contact technical support and inform them of
the error message displayed.
Empty Cell control 1 out of range
Empty Cell control 2 out of range
Empty weight out of range.
Genuine Aquamatic Validation Error. Please contact customer service.
Sample magnitude is not stable
Sample phase is not stable
Unable to Initialize ACC
Unknown Error
Validation Error