F3 Corpus
Event
Possible cause
Remedy
The wheelchair stops while being driven.
The cable connection to the control
panel has come loose.
Insert the cable in the control panel.
Main circuit breaker tripped.
See page 246.
The wheelchair can only be driven at
reduced speed.
Seat lift or seat tilt raised too high.
Applies only to a powered seat lift and
seat tilt.
Lower the seat lift or seat tilt.
The wheelchair cannot be charged.
Main circuit breaker switched to off
position after e.g. battery replacement.
See page 246.
The charging circuit breaker has tripped.
Wait five minutes, the circuit breaker will
automatically reset.
8.2
Diagnostics R-Net LCD control
panel
8.2.1
General
When an error or a fault occurs in the wheelchair’s electronics,
information about it is displayed in the control panel display. This
information can then be used to diagnose where the error, or fault,
occurred and its cause.
261
Summary of Contents for F3 Corpus
Page 1: ...User manual British English F3 Corpus...
Page 2: ......
Page 4: ...Text Produced and published by Permobil Edition 12 Date 2019 09 03 Order no 334968 eng UK...
Page 6: ......
Page 10: ......
Page 16: ...F3 Corpus Important information 16...
Page 36: ...F3 Corpus Safety instructions 36...
Page 132: ...F3 Corpus Preparing for use 4 4 2 Air transportation 194 4 4 3 Rail transportation 195 132...
Page 186: ...F3 Corpus Preparing for use Figure 165 Front tie down eyes Figure 166 Rear tie down eyes 186...
Page 196: ...F3 Corpus Preparing for use 196...
Page 229: ...F3 Corpus Maintenance and repairs D Rim outer section E Hub cap F Wheel accent covers 229...
Page 250: ...F3 Corpus Maintenance and repairs 250...
Page 251: ...F3 Corpus Approved locking systems 6 Approved locking systems Dahl Docking System MKII 251...
Page 252: ...F3 Corpus Approved locking systems 252...
Page 258: ...F3 Corpus Restrictions and limitations 258...
Page 274: ...A 334968 eng UK 7 330818 318992...