background image

The Bead Ruptor Elite

TM

 Bead Mill consists of the following:

Description

Quantity

Bead Ruptor Elite

TM

 Bead Mill Assembly

1

Universal Finger Plate

1

Power Cord, US, EU, UK, AU

1 Each

Spare Fuse

1

User Manual

1

Overview

Voltage

1. Lid

2. Drive Hub

3. Tube Carriage

4. Lid Latch

1.

2.

3.

4.

5.

6.

7.

8.

5. Touch Screen

6. 120V/230V Fuse

7. ON/Off Switch

8. Power Cord Port

220 volts

- Remove Warning stickers from the back of the unit and check that the voltage matches the 

  voltage delivered by your electric power supply.

CAUTION: DO NOT connect the unit to the main power supply until correct voltage is selected.
- Plug the unit into power supply using the supplied power cord.
CAUTION: If the voltage supplied to the unit is higher than what is indicated on the voltage 

switch, the electronics can be irreversibly damaged, voiding the warranty.

BEAD RUPTOR ELITE

Bead Mill Homogenizer

15 = 115 - 120V
30 = 220 - 230V

115V

115V

Summary of Contents for OMNI Bead Ruptor Elite

Page 1: ...Bead Ruptor EliteTM BEAD RUPTOR ELITE Bead Mill Homogenizer Bead Mill Homogenizer User Manual ...

Page 2: ... OF MERCHANTABILITY OR OTHERWISE EXPRESSED OR IMPLIED IN FACT OR BY LAW AND STATE OUR ENTIRE AND EXCLUSIVE LIABILITY AND BUYERS EXCLUSIVE REMEDY FOR ANY CLAIM OF DAMAGES IN CONNECTION WITH THE SALE OR FURNISHING OF GOODS OR PARTS THEIR DESIGN SUITABILITY FOR USE INSTALLATION OR OPERATION OMNI INTERNATIONAL WILL IN NO EVENT BE LIABLE FOR ANY SPECIAL OR CONSEQUENTIAL DAMAGES WHATSOEVER AND THEIR LIA...

Page 3: ...lbs The dimensions of the system are withlidclosed 38 1x43 18x33cm 15 x17 x13 withlidOpen 39 37x44 45x61 6cm 15 5 x17 5 x24 25 Operating Environment 4 C to 40 C 39 F to 104 F Humidity 5 to 95 RH Ambient Air The system must be located in an area where the ambient air is clean No emission of solid particles or smoke in the air by adjacent equipment is allowed The level of dust should be comparable t...

Page 4: ...impaired FCC This device complies with part 15 of the FCC United States Federal Communications Commission Rules Operation is subject to the following two conditions This device may not cause harmful interference and This device must accept any interference received including interference that may cause undesired operation CE This device complies with all CE rules and requirements Changes or modifi...

Page 5: ...d Port 220 volts Remove Warning stickers from the back of the unit and check that the voltage matches the voltage delivered by your electric power supply CAUTION DO NOT connect the unit to the main power supply until correct voltage is selected Plug the unit into power supply using the supplied power cord CAUTION If the voltage supplied to the unit is higher than what is indicated on the voltage s...

Page 6: ...must be spaced evenly Place Tube Carriage sold separately on Drive Hub Place Finger Plate on Drive Hub Screw down to secure tubes Close the lid until the locking handle clicks into place CAUTION The unit will not run without the lid fully closed WARNING Do not run the Bead Ruptor EliteTM Bead Mill without a tube carriage installed 2 4 5 1 3 ...

Page 7: ...on Getting Started Language Setting 1 Menu 2 Home 3 Search 4 Configure 1 2 3 4 Select Menu Select Configure Select Language v wTo navigate through menu Choose from French English German Italian or Spanish Select Home ...

Page 8: ...o Create new program 1 Menu 2 Program Name 3 Tube Setting 4 Speed 5 Time 6 Processing chamber temperature 7 Run Pause 8 Tube Speed Limit 9 Number of Cycles 10 Time interval or Dwell between cycles 1 2 3 4 5 6 7 8 9 10 00 30 00 30 02 4 50 Max Tube Speed 08 46 2 30 ml Tube Bacteria 30 mL 28 C Ready ...

Page 9: ...ram with new name Delete program Edit Title Change program name Parameter Operating Range S Speed Energy From 0 8 m s to 8 m s T Time From 0 01 to 9 59 in increments of 1 second C Number of Cycles 1 to 10 D Dwell Pause between Runs 0 00 to 9 59 Run time indicator Pause Run button Cycle indicator Stop run button Bacteria 30 mL 28 C 02 35 ...

Page 10: ...n Faulty display screen 1 Turn off the unit 2 Contact technical support One or several tubes are not sealed tightly The cap is not properly sealed or the tube is faulty If dangerous or potentially dangerous sample is contained in the tube apply the proper decontamination procedure Unit powers on with a loud BANG Motor control device damaged due to incorrect voltage selection Return unit to OMNI In...

Page 11: ......

Page 12: ...PN 03 241E Rev C 935 Cobb Place Blvd Kennesaw GA 30144 800 776 4431 770 421 0058 www omni inc com ...

Reviews: