Installation
For all products, check the unassembled parts against the Packing List for the respective product to ensure that all
items have been included.
Tools needed (included)
• 10 mm wrench for feet adjustment
• 3 mm hex key for lid replacement
Tools needed (not included)
• 16 mm wrench for attaching tubing to circulating water bath
• screw driver (slot head) for attaching tubing to circulating water bath
Figure 6: Installation kit contents
For IQ/OQ, the FR-9120/FR-9400/FR-9360 RESOLVE Hemoglobin kit and the 3412-0010 Resolve Hb FASC Control kit or
equivalent controls are needed (not included). Use the same kit and gel that you are planning to use for the samples.
As the Migele Gel Electrophoresis Unit is intended to control the gel processing condition, avoid warm exhaust air
from other instruments when placing the instrument. Leave sufficient space behind the instrument for tubing and
ventilation.
Remove the transport protection from the top of the lid and the protective plastic sheet from the cooling plate. For
more details, see the pictures below.
2118-0010 - Migele
®
Gel Electrophoresis Unit User Manual
Installation
11