Temperature Display
g.
Select which temperature will be displayed on the panel:Block, Probe 1 or Probe 2. The only
selected temprature is displayed in real time.
h.
Offset
: This function is used to adjust temperature.
It is only used for manfacturing
process, so do not modify the values.
10.
After completing the parameter setup, select
Apply
and
OK
. Then the LED of ON-LINE is turned on,
it will start heating up or cooling down to the set temperature in the Setup Multi Cell Peltier tab.
11.
Insert blank solutions into the both reference and sample cell holder. Select
Baseline (or Zero)
and
then baseline (or zero) will be measured when it reaches the set Temperature.
NOTE:
To monitor the probe temperature, the probes should be immersed in the sample, or to use the block
temperature for monitoring, the cell lid should be closed tightly.