09931463A
26
b.
Using Cells
: Show the cell positions which are selected for measurement. Remove the cells by
and switch positions by
and
keys.
c.
Multi Zero:
If checked, zero will be measured all the selected positions. If not, zero will be only
measured at the 1
st
cell position among the using cells.
NOTE:
To prevent inadequate baseline (zero) measurement because of the difference in transmittance among
cuvettes or any other reasons, Multi Baseline (Multi Zero) measurement is recommended.
d.
Temperature (°C)
: Enter the returning temperature after the experiment is finished.
e.
Monitor:
Select temperature that used to monitor during measurement. Options include: Block,
Probe1 or Probe2.
f.
Stabilizing Time (S)
: Set the stabilization time after reaching the set temperature.
g.
X Axis Value
: Select the displayed x-axis value of the graph; probe temperature or block
temperature.
h.
Start (°C)
: Enter the start temperature for the measurement.
i.
End (°C)
: Enter the end temperature for the measurement.
j.
Interval (°C)
: Enter the measurement interval temperature. For instance, if you enter 5°C, the
sample will be measured from start temperature to end temperature every 5°C.
k.
Rate (°C/min)
: Enter the heating (or cooling) temperature rate in each temperature range.
NOTE:
The rate has to be set under or equal to interval value.
l.
Hold (min)
Enter the holding time. For example, if you set 1 min, measurement will start one minute after
temperature reaches at the set point and the set stabilizing time has elapsed.
m.
Error Range (°C/min)
It shows the temperature tolerance between the sampling and monitored temperature. The
measurement will start when the temperature tolerance reaches within the set Error Range.
Example: Set the starting temperature at 20°C and error range with 1°C then, the
measurement will start when the temperature reaches between 19°C and 21°C.
▶
Temperature Display
n.
Select which temperature will be displayed on the panel:Block, Probe 1 or Probe 2.
The selected temperature is only displayedin real time.
o.
Offset
:
This function is used to adjust temperature.
It is only used for manufacturing process, so
do not modify the values.