Comments:
PerkinElmer LAS Representative:
Date:
Reviewed By:
Date:
Flexar Analytical / FX-10 UHPLC Pump
IQ/OQ Rev. D June 2009
2.4.7
Test Results
PARAMETER
SPECIFICATION
TEST
RESULT
PASS/FAIL
Composition Accuracy
Within ± 1.0% absolute @ 5 mL/min
Channel A and B
% Solvent B
10%
49%
51%
90%
49 / 51 Delta Range
Between 1.5% and 2.5%
Channel C and D
% Solvent D
10%
90%
Comments
:
2.5 Operational Qualification Verification
The Flexar Analytical / FX-10 UHPLC Pump, Serial Number Template-SN,
operates
and performs according to the stated Operational Qualification, and all appropriate
forms and documents supporting the system have been filled out and are available.
Verified: