Flexar Analytical / FX-10 UHPLC Pump
IQ/OQ Rev. D June 2009
Pressure Pulsations
Pulsations (PSI) at 1 ml/min solvent flow and 1000 PSI backpressure
(approx)
Flow Accuracy
For Analytical pumps, solvent flow as measured at 1ml and 3ml per
minute.
For the FX-10 UHPLC pumps, solvent flow as measured at 0.3ml and
0.9ml per minute.
Compositional Accuracy
For Binary / Quaternary Analytical pumps, accuracy as measured by
relative UV absorbance (265 nm) at programmed proportions of 0.2%
Acetone/Water and Water alone.
PerkinElmer assumes no responsibility for test results.
2.4
Parameter Testing:
2.4.1 Pressure Pulsations Test
Test Conditions