58 . Frontier IR Single-range Spectrometers User's Guide
Renewing the instrument desiccant
CAUTION
Old, used desiccant releases moisture and can cause catastrophic failure
of KBr optics.
Do not use damaged packs of desiccant. Make sure that the packs you
use have not been left in contact with the air.
You can purchase disposable desiccant kits from PerkinElmer (part number N0171159). A kit
contains two packs of desiccant, and three kits are required.
1.
Inspect the humidity indicator card in the plastic bags in which the spare desiccant
packs are packed. If the card indicates humidity in the bag, discard the desiccant pack.
2.
Remove the sample area cover, if fitted, by opening the cover, pressing the clip and
pulling the cover vertically to remove (Figure 25).
Figure 25 Removing the sample compartment cover
3.
Undo the two captive screws securing the desiccant cover (Figure 26).
Figure 26 Captive screws securing the desiccant cover
Screw
Screw
Sample cover
release clip
Summary of Contents for Frontier FT-IR
Page 1: ...FRONTIER IR SINGLE RANGE SYSTEMS User s Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 10: ...10 Frontier IR Single range Spectrometers User s Guide ...
Page 11: ...Warnings and Safety Information ...
Page 23: ...An Overview of the Frontier IR Single Range Spectrometers ...
Page 32: ...32 Frontier IR Single range Spectrometers User s Guide ...
Page 33: ...Unpacking and Installation ...
Page 43: ...Using the Spectrometer with Spectrum ...
Page 49: ...Using the Spectrometer with Spectrum 49 Figure 20 Atmospheric suppression ...
Page 53: ...Routine Maintenance ...
Page 68: ...68 Frontier IR Single range Spectrometers User s Guide ...
Page 69: ...Advanced Maintenance ...
Page 88: ...88 Frontier IR Single range Spectrometers User s Guide ...
Page 89: ...Appendices ...
Page 100: ...100 Frontier IR Single range Spectrometers User s Guide ...