36 . Frontier IR Single-range Spectrometers User's Guide
2.
Carefully remove the instrument from the shipping container, but not from the bag in
which it was shipped.
CAUTION
The spectrometer must be allowed to reach the temperature of its
surroundings before it is removed from the bag. This means leaving it
overnight if it has been moved from a cold area, and at least 4 hours
after removal from the shipping container.
3.
When the spectrometer has been allowed to warm to the temperature of its
surroundings, remove it from the bag and place it on the bench where it is to be used.
Ensure that you can reach the rear of the spectrometer to connect the cables.
NOTE:
Any accessories will be shipped in separate boxes.