30 . DSC 4000 Installation and Hardware Guide
CAUTION
ATTENTION
Care must be taken when handling the outer furnace lid so its ceramic
lining is not damaged.
Des précautions doivent être prises lors de la manipulation du couvercle
du four extérieur, de sorte que sa doublure en céramique n'est pas
endommagée..
•
Avoid contamination of the furnace, use high-purity purge gases to purge any reaction
products out of the furnace, and heat the furnace to temperatures above 430 °C ONLY
when absolutely necessary.
•
Check that the cooling liquid supply is circulating before starting a run.
•
Ensure the purge gases are set to the recommended flow rates. A rate that is too high
will disturb the inner furnace lid.
•
Do not operate the instrument in a cold room. The ambient temperature and the
temperature of the instrument should be between 10 °C and 30 °C.
•
Never open the furnace when it is at an elevated temperature.
Important Specifications
Maximum power consumption
430 VA
Supply voltage
100–120 V or 220–240 V
Supply voltage frequency
50–60 Hz
Fuses
6.3 AT or 3.15 AT
Safe temperature range
10–40 °C
Dimensions L x H x D
38 x 17 x 41 cm (H 29 cm with autosampler)
Weight
17 kg (19 kg with autosampler)
Summary of Contents for DSC 4000
Page 1: ...DSC 4000 Installation and Hardware Guide THERMAL ANALYSIS...
Page 5: ...Introduction...
Page 7: ...Safety and Regulatory Information...
Page 17: ...Safety and Regulatory Information 17...
Page 18: ...Prepare the Laboratory...
Page 21: ...Prepare the Laboratory 21...
Page 22: ......
Page 23: ...Install Multiple Analyzers...
Page 26: ...26 DSC 4000 Installation and Hardware Guide...
Page 27: ...Install a DSC 4000...
Page 52: ......
Page 53: ...DSC 4000 Hardware...
Page 65: ...AS 6000 Autosampler...