14
United States (FCC)
This instrument is classified as a digital device used exclusively as industrial,
commercial, or medical equipment. It is exempt from the technical standards
specified in Part 15 of FCC Rules and Regulations, based on Section 15.103 (c).
Environment
Operating Conditions
Pollution Degree
This equipment will operate safely in environments that contain non-conductive
foreign matter up to Pollution Degree 2 in EN 61010–1.
Recommended operating conditions:
• Indoors.
• The location must be free of smoke, dust, corrosive fumes, direct sunlight
and excessive vibration.
• Ambient tempe15
C to +35
C (+59
F to +95
F), with a
maximum change not exceeding 2.8
C (5
F) per hour. For optimum
instrument performance, the room temperature should be controlled at 20°
± 2 °C (68° ± 2 °F).
• Relative humidity 20% to 80%, without condensation.
• Altitude in the range -400 m to 2,000 m (-1,312 ft. to 6,562 ft).
Safe operating conditions:
• Indoors
Explosive Atmosphere
This instrument is not designed for operation in an explosive atmosphere.
Avertissement : atmosphère explosive
Cet instrument n'est pas conçu pour fonctionner dans une atmosphère
explosive.
!
Warning
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...