Safety/Regulatory Information
1
-24
0993-5255
Safety practices for FIAS and mercury hydride systems
These systems are offered as options for the spectrometer.
Safety checks
Before you start an analysis, make sure that:
–
The laboratory fume ventilation system is operating;
–
The tubing in the fluid system is not constricted since this could lead to a
pressure build-up in the system;
–
The drain tubing is inserted in the waste container and that this container
is empty.
If you are using a flame to heat the QTA-cell or if you intend to perform
flame-FIAS analyses, refer to the information under
Safety Practices for
Flame Atomization
before you ignite the flame.