Row Count
Enter the value for the number of rows.
Start X, Start Y
Enter the x and y distance of the A1 well to the upper right corner of
the plate.
Interval X,
Interval Y
Enter the distance between the centers of two neighboring wells in
the x and y directions.
Radius (mm)
Enter the value for the radius of the wells in the plate.
Reload
Open the previously saved parameters.
Save
Save the setting parameters.
Manual Alignment Refer to the Manual Alignment chapter.
Show Dimensions
If you check Show Dimensions, it shows the range of each
parameter.
5.
Click Save.
6.
Click the Lamp On button in the Operation to turn on the lamp. Enter the A1 and click Move
to button.
7.
If the A1 position is not correct, Go to the next Manual Alignment section.
NOTE:
The Start X value and the Start Y value may differ because the starting position differs depending on
the accessories and equipment.